Table 2.

Limit of detection and precision of chemiluminescence immunoassays.

Validation Parameter CL-S-ELISA 1 CL-C-ELISA 1
Limit of detection (ng/mL) 0.5 5
Intra-assay precision (CV% 2) 1.5–4.3 5.1–7.5
Inter-assay precision (CV%) 2.7–7.7 6.2–13.5

1 The limit of detection, intra-assay precision, and inter-assay precision were calculated using trout samples non-spiked and spiked with A. simplex antigen, which were subsequently heated for 60 min at 100 °C; 2 CV%, percent coefficient of variation.