Table 1.

Comparison of RT-PCR and rapid antigen detection kits for the limit of detection of SARS−COV-2 virus.

Dilutionb Test resultsa
RAD kit
RT-PCRc
BIOCREDIT Coris NADAL Standard Q
10−1 POS POS POS POS 15.15
10−2 POS POS POS POS 18.44
10−3 NEG NEG NEG POS 21.54
10−4 NEG NEG NEG POS 24.64
10−5 ND ND ND POS 28.67
10−6 ND ND ND NEG 31.90
10−7 ND ND ND ND 34.87
10−8 ND ND ND ND NEG
a

POS, positive; NEG, negative; ND, not done.

b

Serial tenfold dilution of the respiratory specimen, NPA & TS, obtained from the Hong Kong COVID-19 patient, hCoV-19/Hong Kong/VM20030475/2020.

c

RT-PCR were tested twice with identical results. The Ct values shown were the mean of both runs.