Table 1.
Comparison of RT-PCR and rapid antigen detection kits for the limit of detection of SARS−COV-2 virus.
Dilutionb | Test resultsa |
||||
---|---|---|---|---|---|
RAD kit |
RT-PCRc | ||||
BIOCREDIT | Coris | NADAL | Standard Q | ||
10−1 | POS | POS | POS | POS | 15.15 |
10−2 | POS | POS | POS | POS | 18.44 |
10−3 | NEG | NEG | NEG | POS | 21.54 |
10−4 | NEG | NEG | NEG | POS | 24.64 |
10−5 | ND | ND | ND | POS | 28.67 |
10−6 | ND | ND | ND | NEG | 31.90 |
10−7 | ND | ND | ND | ND | 34.87 |
10−8 | ND | ND | ND | ND | NEG |
POS, positive; NEG, negative; ND, not done.
Serial tenfold dilution of the respiratory specimen, NPA & TS, obtained from the Hong Kong COVID-19 patient, hCoV-19/Hong Kong/VM20030475/2020.
RT-PCR were tested twice with identical results. The Ct values shown were the mean of both runs.