Table 1.
MRI scan parameters for T1-weighted, T2 measurements and STEAM diffusion scans
T1 weighted | T2 | Diffusion | |
---|---|---|---|
Imaging sequence | Turbo spin-echo (TSE) | Multi-echo, spin-echo (MESE) | STEAM-EPI |
TR: repetition time (ms) | 697 | 1500 | 6300 |
TE: echo time(s) (ms) | 9.1 | 9.6:9.4:153.6 (16 echos) | 42.2 |
Field of view (mm) | 300 × 300 | 300 × 300 | 300 × 300 |
Slice thickness (mm) | 5 | 5 | 5 |
Fat suppression | STIR | SPAIR | SPAIR |
Acquisition matrix | 256 × 256 | 256 × 256 | 256 × 256 |
Number of slices | 60 | 5 | 5 |
Number of averages | 1 | 1 | 8 |
Receiver bandwidth (Hz/pixel) | 222 | 510 | 1502 |
Flip angle (°) | 90 | 15 | - |
GRAPPA | - | - | 2 |
b values (s/mm2) | - | - | 0, 500 |
Directions | - | - | 6 |
Mixing time (Tm) (ms) | - | - | 980 |
Diffusion time (∆) (ms) | - | - | 1000 |
Acquisition time (min:s) | 2:19 | 2:05 | 6:12 |