Skip to main content
. 2020 May 20;27(5):1728–1738. doi: 10.1007/s12350-020-02200-6

Table 2.

82Rb generator daily QA activity (n = 9 each)

Generator system Integral (yield) Maximum (steady-state)
GBq (mCi) on first day of use GBq (mCi) on last day of use GBq (mCi) on first day of use GBq (mCi) on last day of use
RUBY-FILL®*

3.04  ±  0.17

(82.2  ±  4.6)

on day 4

0.80  ±  0.01

(21.7  ±  0.1)

on day 56

2.31  ±  0.13

(62.6  ±  3.5)

on day 4

0.61  ±  0.01

(16.5  ±  0.08)

on day 56

CardioGen-82®**

3.03  ±  0.21

(82.0  ±  5.6)

on day 0

1.07  ±  0.06

(28.9  ±  1.6)

on day 42

2.31  ±  0.16

(62.4  ±  4.3)

on day 0

0.81  ±  0.04

(22.0  ±  1.2)

on day 42

Values are mean ± standard deviation. mCi values = GBq × 27 mCi/GBq

* RUBY-FILL Maximum values = Integral values ÷ 1.314 correction-factor

**CardioGen-82 Integral values = Maximum values × 1.314 correction-factor