Repeatability and non-linearity error (%) for AMI air-pack sensors of batch 1 and 2.

Sensor Nonlinearity Error (%) Repeatability Error (%) Hysteresis Error (%) Accuracy (mmHg)
Batch 1
Sensor 1 1.22 3.62 2.37 2.67
Sensor 2 0.52 1.48 1.60 1.41
Sensor 3 0.64 2.60 3.39 2.65
Sensor 4 0.31 2.00 3.18 2.35
Batch 2
Sensor 1 1.17 2.08 2.69 2.18
Sensor 2 0.97 2.08 0.91 1.53
Sensor 3 0.88 3.07 1.39 2.19
Sensor 4 1.27 2.67 1.91 2.19