. 2020 Sep 28;12(10):2786. doi: 10.3390/cancers12102786

Table 1.

Technical validation of the C4G assay.

Technical Validation Step Results
LLMR (uncorrected for pre-dilution) 2.3 ng/mL
ULMR (uncorrected for pre-dilution) 788 ng/mL
Calc. midpoint (IC50) 8.8 ng/mL
Intra-assay variation 6%
Inter-assay variation 8%
Dilution recovery of serum 94%
Dilution recovery of EDTA plasma 106%
Spiking recovery of serum in serum 102%
Analyte recovery, 4 freeze/thaw cycles 96%
Analyte recovery 24h, 4 °C/20 °C 109%/106%
Analyte recovery 48h, 4 °C/20 °C 122%/109%
Interference test
Biotin recovery, low/high 94%/71%
Lipemia recovery, low/high 111%/102%
Hemoglobin recovery, low/high 99%/92%

LLMR: lower limit of measurement range; ULMR: upper limit of measurement range. Percentages are reported as mean.