Table 1.

Comparison of reported Lyme disease biosensors. Bold font represents biosensors tested specifically against early Lyme disease samples.

Method Sample Target Sensitivity or LOD Specificity Time Ref.
Microfluidics Serum Antibodies 84% 92% 15 min [32]
Serum Antibodies 80%/85% 95% - [33]
FET Spiked Buffer Antigen 1 ng/mL High Minutes [35]
Spiked Buffer Antigen 2–500 pg/mL High - [37]
LFA Serum Antibodies 64.2%/80% 89.5%/96.5% 3–15 min [39]
VFA Serum Antibodies 162.2–1046 ng/mL - 20 min [41]
Serum Antibodies 85.7% 96.3% 15 min [42]
SPR Serum Antibodies 81%/92% 86%/82% A few min [44]
Biochip Serum Antibodies 0.39–0.78 µg/mL - - [46]
Serum Antibodies 90% 100% - [47]

LOD = Limit of Detection, FET = Field Effect Transistor, LFA = Lateral Flow Assay, VFA = Vertical Flow Assay, SPR = Surface Plasmon Resonance.