Comparison of the available tracers for quantitative perfusion PET

[15O]water [13N]NH3 82Rb [18F]flurpiridaz
Radionuclide half-life 122 s 9.96 min 75 s 109.8 min
Availability On-site cyclotron On-site cyclotron Generator Cyclotron (possible shipping)
Mean positron range in water (mm) [20] 2.5 1.5 5.9 0.6
Relationship with MBF Ideal (freely diffusible) High extraction fraction Non-linear extraction fraction High extraction fraction
Image quality Parametric MBF images only Good to high Fair to good Very high
Gated imaging Possible from first pass (blood pool) High quality Good quality High quality
Time schedule Very tight Interval between rest and stress injections Very tight Separate stress and rest injections
Previous experience Used mainly in research setting and with hybrid imaging Widely used qualitatively and quantitatively Widely used qualitatively and quantitatively Presently ongoing phase III trial

MBF = myocardial blood flow