Table 1.  . Summary of validation outcomes: intra and inter day accuracies and precisions for vancomycin in human whole blood volumetric absorptive microsampling.

Conc. Mean ± SD (μg/ml) CV (%) Accuracy (%) Mean ± SD (μg/ml) CV (%) Accuracy (%)
  Day 1 (n = 6 replicates) Day 2 (n = 6 replicates)
1.00 1.12 ± 0.129 11.5 112 0.959 ± 0.114 11.9 95.9
2.50 2.68 ± 0.288 9.91 107 2.48 ± 0.085 3.45 99.1
40.0 39.8 ± 0.955 2.40 100 38.4 ± 1.73 4.50 96.1
80.0 84.3 ± 10.0 11.9 105 88.5 ± 9.80 11.1 111
  Day 3 (n = 6 replicates) Inter day (3 days, n = 18 replicates)
1.00 0.903 ± 0.082 9.09 90.3 0.993 ± 0.139 14.0 99.3
2.50 2.59 ± 0.253 9.74 104 2.59 ± 0.225 8.67 104
40.0 37.6 ± 1.57 4.16 94.0 38.6 ± 1.65 4.28 96.5
80.0 76.4 ± 3.76 4.92 95.4 83.0 ± 9.42 11.3 104