Table 3.

Metal artifact size and reduction in size among sequences

Imaging method Average artifact size (mm)* Standard deviation (mm) Average artifact reduction (mm) % “best” image quality Average Quality Rating
Standard SSFP cine 24 9.6 4% 2.2
Optimized SSFP cine 22 9 1.5 12% 2.4
Optimized GRE cine 19.4 8.1 4.6 84% 2.7
Standard TSE black blood 15.4 5.4 25% 2.8
Optimized TSE black blood 13.6 1.6 1.6 75% 3.1

GRE gradient recalled echo, SSFP steady-state free precession, TSE turbo spin echo

*

Average artifact size is the average of two orthogonal measurements of the artifact in millimeters