Table I. Key Components of the PTMS.

Component Assignment Nominal Rating Part Number Manufacturer
T Step-up transformer Output: 1 kV,10 A Class-E insulation Custom manufactured Eastern Transformers, UK
Rectifier Full bridge diode rectifier 1200 V, Ultrafast recovery diode IFRM a = 600 A STTH9012TV1 STMicroelectronics
RDC Charging resistor 2x 47 Ω, thick-film on steel 2.5 kV, 1.5 kW (parallel) WDBR1-47RKT TT ELECTRONICS
CDC Energy storage 10000 μF, 500 VDC Aluminum capacitor ALS70A103NT500 KEMET Electronics
Q1-Q4 IGBT power switch 1.2 kV ICRM b = 1.8 kA QG = 3.45 µC SEMiX603GB12E4p Semikron
D Free-wheeling diode (included in the IGBT module) 1.2 kV IFRM c = 1.8 kA
C Snubber capacitor 2 µF, Polypropylene film capacitor 205PPA122K Illinois Capacitor
R Snubber resistor 1Ω, non-Inductive film resistor AP821 1R F Arcol
Scale 2+ driver core Gate driver core VGEon= 15 V, VGEoff= -8 V, IG = 6A. 2SC0106T2A1-12 Power Integrations
RE Emitter resistor 500 mΩ, AP821 R5 J Arcol
L Stimulation coil 15.5 µH D70 Remote Coil Magstim
ctrl Digital controller PWM generation resolution: 10 ns MicroLabBox, includes Power PC Dual Core 2 GHz processor, DS1202, DS1302 I/O dSPACE
a)

Repetitive peak forward current, tp = 5 µs, F= 5 kHz square

b)

Peak current value at collector output during pulse operation.

C)

Repetitive peak forward current of the free-wheeling diode.