Troubleshooting.
Step | Problem | Possible reason | Solution |
---|---|---|---|
7 | Fibers are not straight and are curly Fibers are clumped |
Fibers have been shaved too much Fibers have not been separated well enough |
Try a more upright angle of the scalpel blade in order to cut individual fibers off without shaving them or dragging the blade along the fibre Use a narrower angle in order to separate individual fibers while cutting Press down on the suture to splay the individual fibers before cutting |
10 | Cells are not coating the fibers and are clumped in the middle Microscaffold EBs are very round with fibers inside the sphere |
Fibers are too long Fibers are too short |
Make sure to cut fibers about 1mm in length Same as above |
12 | EBs are very dark | Lack of neuroectoderm EBs are too large |
Check the quality of starting PSCs. PSCs grown in StemFlex or E8 tend to be more efficient at neural induction. Try the commercially optimized kit. Small molecules (dual SMAD inhibitors) may be needed. Check the accuracy of the cell counting |
31 | A lot of outgrowth of non-neural identities | Failed neural induction Too much neural crest |
Same as above Try again, it may be an issue with the batch Try the commercially optimized kit |
39 | Buds remain small and underdeveloped Lack of formation of a cortical plate |
Shaker speed is not optimal Tissue culture vessel or volume is not optimal Shaker speed or vessel is not optimal Matrigel is not fully dissolved |
Be sure to adjust shaker speed depending on orbit diameter Use a 60mm dish with the shaker speed at 57rpm. A smaller or larger dish will result in non-optimal centrifugal forces Same as above Be sure to add Matrigel to cold media just before feeding to avoid Matrigel precipitation |
51 | The organoid tissue detaches from the agarose matrix during sectioning | Leftover Matrigel or media around the organoid can cause the tissue to detach | Remove all matrigel from the surface of the organoid prior to embedding Rinse the organoid in HBSS and incubate in warm agarose for a few minutes before placing in embedding mold and cooling |
51 | The organoid is damaged during sectioning | Sectioning speed and/or vibration frequency may be too high Necrosis has already developed in the core The organoid is too large |
Decrease the sectioning speed and the frequency of vibration Section at earlier stages Section at earlier stages and do not let the organoids fuse |
52 | Difficulties handling sections and maintaining tissue integrity | Too much or too little agarose left around the organoid The bristles of the brush come into contact with the tissue |
Cut the agarose around the organoid leaving ~2-mm space all around the organoid Orient the block on the specimen plate so as to minimize the travel distance of the blade Only manipulate the agarose surrounding the organoid section Use two scalpels to manipulate the sections |
58 | Poor ALI-CO health after sectioning Contamination has spread across multiple inserts |
Too slow organoid transfer to agarose Too long incubation on ice The antibiotic-antimycotic solution used became spoiled The same tip was used to remove spent medium from multiple wells The repeated pipette combitip came into contact with a contaminated surface |
Embed fewer organoids at one time and work quickly Embed fewer organoids at one time After thawing an aliquot do not refreeze and discard Thaw the antibiotic solution at 4 °C overnight For each well, use a new 1 ml tip to remove the medium Do not touch the side of the wells containing the ALI-COs or any other potentially non-sterile surface with the combitip |
Box 3,step 13 | Problems loading or expelling the micropipette | Bubbles in the micropipette that interfere with the vacuum | Use a 10 ml syringe to push the air bubbles out of the micropipette. Wash with PBS and ensure there are no further bubbles. |