Skip to main content
. Author manuscript; available in PMC: 2021 Aug 1.
Published in final edited form as: Nat Protoc. 2020 Dec 16;16(2):579–602. doi: 10.1038/s41596-020-00433-w

Troubleshooting.

Step Problem Possible reason Solution
7 Fibers are not straight and are curly

Fibers are clumped
Fibers have been shaved too much

Fibers have not been separated well enough
Try a more upright angle of the scalpel blade in order to cut individual fibers off without shaving them or dragging the blade along the fibre
Use a narrower angle in order to separate individual fibers while cutting Press down on the suture to splay the individual fibers before cutting
10 Cells are not coating the fibers and are clumped in the middle

Microscaffold EBs are very round with fibers inside the sphere
Fibers are too long

Fibers are too short
Make sure to cut fibers about 1mm in length

Same as above
12 EBs are very dark Lack of neuroectoderm

EBs are too large
Check the quality of starting PSCs. PSCs grown in StemFlex or E8 tend to be more efficient at neural induction. Try the commercially optimized kit. Small molecules (dual SMAD inhibitors) may be needed.

Check the accuracy of the cell counting
31 A lot of outgrowth of non-neural identities Failed neural induction

Too much neural crest
Same as above

Try again, it may be an issue with the batch
Try the commercially optimized kit
39 Buds remain small and underdeveloped


Lack of formation of a cortical plate
Shaker speed is not optimal

Tissue culture vessel or volume is not optimal

Shaker speed or vessel is not optimal

Matrigel is not fully dissolved
Be sure to adjust shaker speed depending on orbit diameter

Use a 60mm dish with the shaker speed at 57rpm. A smaller or larger dish will result in non-optimal centrifugal forces

Same as above

Be sure to add Matrigel to cold media just before feeding to avoid Matrigel precipitation
51 The organoid tissue detaches from the agarose matrix during sectioning Leftover Matrigel or media around the organoid can cause the tissue to detach Remove all matrigel from the surface of the organoid prior to embedding

Rinse the organoid in HBSS and incubate in warm agarose for a few minutes before placing in embedding mold and cooling
51 The organoid is damaged during sectioning Sectioning speed and/or vibration frequency may be too high

Necrosis has already developed in the core

The organoid is too large
Decrease the sectioning speed and the frequency of vibration

Section at earlier stages

Section at earlier stages and do not let the organoids fuse
52 Difficulties handling sections and maintaining tissue integrity Too much or too little agarose left around the organoid

The bristles of the brush come into contact with the tissue
Cut the agarose around the organoid leaving ~2-mm space all around the organoid

Orient the block on the specimen plate so as to minimize the travel distance of the blade
Only manipulate the agarose surrounding the organoid section Use two scalpels to manipulate the sections
58 Poor ALI-CO health after sectioning


Contamination has spread across multiple inserts
Too slow organoid transfer to agarose

Too long incubation on ice

The antibiotic-antimycotic solution used became spoiled

The same tip was used to remove spent medium from multiple wells

The repeated pipette combitip came into contact with a contaminated surface
Embed fewer organoids at one time and work quickly

Embed fewer organoids at one time

After thawing an aliquot do not refreeze and discard

Thaw the antibiotic solution at 4 °C overnight

For each well, use a new 1 ml tip to remove the medium


Do not touch the side of the wells containing the ALI-COs or any other potentially non-sterile surface with the combitip
Box 3,step 13 Problems loading or expelling the micropipette Bubbles in the micropipette that interfere with the vacuum Use a 10 ml syringe to push the air bubbles out of the micropipette. Wash with PBS and ensure there are no further bubbles.