Table 9.

Mass spectrometer settings used for nucleosides and caps analyzed herein (CAV: cell acceleration voltage). Settings for quantifiers (quant.) and qualifiers (qual.) are shown.

Compound (quantifier/qualifier) Mode Precursor ion (m/z) Product ion (m/z) Fragmentor voltage(V) CAV (V) Collision energy (V) Retention time (min)
Am (qual.) pos. 282.1 69.1 109 9 25 4.2
Am (quant.) pos. 282.1 136.1 109 9 9 4.2
m6Am (qual.) pos. 296.1 108.0 121 9 68 4.6
m6Am (quant.) pos. 296.1 150.0 121 9 8 4.6
N6pAm (qual.) pos. 320.1 68.8 121 9 25 5.0
N6pAm (quant.) pos. 320.1 174.0 121 9 9 5.0
GpppG (qual.) pos. 789.1 134.7 150 9 121 3.0
GpppG (quant.) pos. 789.1 151.8 150 9 57 3.0
GpppG (qual.) pos. 789.1 248.4 150 9 21 3.0
m7GpppG (qual.) pos. 803.1 151.7 135 9 61 4.3
m7GpppG (qual.) pos. 803.1 165.7 135 9 53 4.3
m7GpppG (quant.) pos. 803.1 248.1 150 9 21 4.3