Table 2. Technical specifications for the DyNeuMo-Mk1 Investigational Research System.

Sensor Characteristics
Inertial sensing 3-axis accelerometer, sensitive to 4 mg activity variations; dynamic range programmable ±2 g to ±16 g; typical sampling rate is 50 Hz
Stimulation Characteristics
Channel Access 8 independent electrodes, typ. arrangement is 2 leads × 4 electrodes
Stimulation overview 2 independent channels, current-controlled, charge-balanced mono-phasic / symmetric biphasic with programmable interphasic delay
Multiplexing Full matrix configuration across electrodes (inc. the case reference)
Stimulation magnitude 0–15 mA (0.05 mA increments) and 0–450 µs pulse width, fractionalized distribution available for guarded cathodes, etc; programmable and independent ramp rates
Recharge characteristics Programmable passive and active recharge, with variable recharge ratio
Stimulation frequency 1–500 Hz frequency stimulation; can go sub-Hz with cycling enabled; independent frequencies available across stimulation channels
Stimulation cycling Adjustable cycle timing for enabling burst stimulation
Stimulation programs Up to 8 independent programs can be configured in the IPG and accessed by algorithms or patient controller
Algorithm Characteristics
Motion Classification Absolute orientation, activity vs non-activity (parameterized), shocks and free-fall detection
Stimulation Control policy Detected classification states mapped to pre-configured stimulation programs with pre-specified transition ramp rate. Two independent stimulation programmes tied to motion states, and a default fall-back.
Risk mitigations Algorithm implementation aligns to 60601-1-10 specifications for physiologic control loops (e.g. limits, alerts, data logs, fallback modes)
Other System Characteristics
Battery capacity, recharge cycle < 2 h recharge with a target recharge interval of 4 days for typical Parkinson’s-like settings
Mechanical Characteristics Cranial-mount, 7.4 cc titanium package, with 2 leads for 4 contacts/lead
Telemetry/External Sensor and Stimulation Synchronisation MICS-band radio, > 1 m distance, with hand-held module
MRI compatibility (in process) MRI conditional imaging for 1.5 T and 3 T imagers
Electrodes Modular design with ability to customize lead length and electrode spacing; currently supports brain stimulation and peripheral extradural electrodes and cuffs
Clinician programmer Standard consumer tablet running Windows 10