Table 5. Accuracy and precision of antimony in human plasma.

Analytical run number 25.0 ng/ml 50.0 ng/ml 500 ng/ml 7500 ng/ml
1 24.8 55.1 486 7420
  26.1 48.6 486 7600
  26.5 47.9 491 7570
  23.2 46 510 7630
  24.2 50.6 495 7610
  26.1 50.6 498 7680
Mean 25.2 49.8 494.0 7585
SD 1.30 3.13 9.05 88.71
%CV 5.16 6.29 1.83 1.17
%Bias 0.80 -0.40 -1.20 1.13
n 6 6 6 6

SD: Standard deviation.