Skip to main content
. Author manuscript; available in PMC: 2022 Dec 16.
Published in final edited form as: Magn Reson Med. 2014 Nov 3;74(5):1423–1434. doi: 10.1002/mrm.25504

FIG. 8.

FIG. 8

Coefficient of variation in B1+ (high coefficient = poor spatial homogeneity) in the idealized coil model plotted against error in SAR estimate for all voxel models. Each blue dot corresponds to a single random RF shim. There is no clear relation between the two variables.