Published in final edited form as: J Geophys Res Atmos. 2018 Feb 23;123(6):3243–3268. doi: 10.1002/2017jd027791

Table 5.

Ozonesonde pump flow efficiencies, ΨP, with 1-σ uncertainties (± ΔΨP). ΨP are taken from the WMO/GAW Report. O3S-DQA panel revised ±ΔΨP values are computed to replace Komhyr, [1986] and Komhyr at al., [1995]± ΔΨP.

Pressure [hPa] Komhyr, [1986] ΨP ± ΔΨP Komhyr et al., [1995] ΨP ± ΔΨP Johnson et al., [2002] 1 ΨP ± ΔΨP O3S-DQA Panel ± ΔΨP

Sfc-100 1.000 1.000 1.000
100 0.993 ± 0.005 0.993 ± 0.005 0.967 ± 0.011 ± 0.010
50 0.982 ± 0.006 0.982 ± 0.005 0.950 ± 0.012 ± 0.012
30 0.978 ± 0.008 0.972 ± 0.008 0.935 ± 0.012 ± 0.014
20 0.969 ± 0.009 0.961 ± 0.012 0.919 ± 0.014 ± 0.017
10 0.948 ± 0.010 0.938 ± 0.023 0.876 ± 0.020 ± 0.022
7 0.935 ± 0.012 0.920 ± 0.024 0.842 ± 0.025 ± 0.028
5 0.916 ± 0.014 0.890 ± 0.025 0.803 ± 0.032 ± 0.037
1

Average of ENSCI-Z model pump efficiency calibrations.