Skip to main content
. Author manuscript; available in PMC: 2020 Nov 5.
Published in final edited form as: Int J Hyperthermia. 2012;28(4):320–336. doi: 10.3109/02656736.2012.680173

Table III.

Summary of in vivo sonication results. Target mean temperature range was 40–41°C. The 3D spatial offset was significantly greater than 0 for the 4mm and 8mm treatment cells (p=0.0002 and p=0.0018, respectively, one-sample t-test), but not for the 12mm treatment cell (p=0.1540, one-sample t-test).

Rabbit Treatment cell size (mm) Mean (°C) SD (°C) T10 (°C) T90 (°C) Diameter of area ≥ 40° C (mm) Mean thermal dose (CEM43) 3D spatial offset (mm)
1 4.0 40.7 1.0 41.9 39.4 4.5 1.9 1.2
2 4.0 40.6 1.0 41.9 39.4 4.5 1.7 1.1
3 4.0 40.4 1.1 41.7 39.0 3.8 2.0 1.1
4 4.0 40.8 1.1 42.2 39.4 4.4 2.7 1.6
5 4.0 40.4 1.0 41.7 39.2 4.2 2.2 0.9
6 4.0 40.6 1.1 41.9 39.2 4.2 1.9 0.6
7 8.0 41.0 1.2 42.5 39.4 9.4 5.5 2.1
8 8.0 40.9 1.2 42.4 39.5 9.3 5.9 2.9
9 (loc. 1) 4.0 40.7 1.2 42.2 39.2 4.5 1.0 1.7
9 (loc. 2) 8.0 41.3 1.4 43.0 39.5 9.1 3.2 1.9
9 (loc. 3) 12.0 41.2 1.5 43.7 39.6 11.6 6.1 4.8
10 (loc. 1) 8.0 40.7 1.1 42.1 39.3 8.1 4.5 2.5
10 (loc. 2) 12.0 40.6 1.1 41.9 39.2 11.47 3.1 2.9

Loc., location