Skip to main content
. Author manuscript; available in PMC: 2020 Nov 5.
Published in final edited form as: Int J Hyperthermia. 2012;28(4):320–336. doi: 10.3109/02656736.2012.680173

Table IV.

Summary of sonication duration in different subtrajectories (as percentages of total treatment time), sonication energies, and sonication efficiencies in vivo. Energies and efficiencies are reported as mean ± SD.

Rabbit Treatment cell size(mm) Target tissue Sonication on (%) 4 mm subtr.(%) 8 mm subtr.(%) 12 mm subtr.(%) Wait subtr.(%) Energy (kJ) Efficiency(cm3/MJ)
1 4.0 Tumour 70 70 - - 30 4.4±0.4 24±0.8
2 4.0 Tumour 60 60 - - 40 3.7±0.4 30±5.2
3 4.0 Tumour 85 85 - - 15 6.2±0.2 14±1.7
4 4.0 Tumour 55 55 - - 45 5.7±0.6 11±1.9
5 4.0 Tumour 70 70 - - 30 5.1±0.7 14±5.1
6 4.0 Tumour 65 65 - - 35 4.4±1.5 29±8.3
7 8.0 Tumour 71 6 65 - 29 7.1±0.4 80±5.0
8 4.0 Muscle 39 39 - - 61 2.5 30
8 8.0 Muscle 52 4 48 - 48 4.5 110
8 12.0 Muscle 53 11 4 38 47 5.6 250
9 8.0 Muscle 64 3 61 - 36 3.1±0.4 160±20.0
10 8.0 Tumour 67 4 63 - 33 5.5±0.4 80±8.4
10 12.0 Tumour 79 12 4 63 21 6.9±0.3 180±7.7

Subtr., subtrajectory.