Table 2.

Diagnostic accuracy of hypertension using the new ambulatory blood pressure monitoring (ABPM) device.

SBP 1 SBP 2 SBP 3 DBP 1 DBP 2 DBP 3 Hypertension
Device Manual Device Manual Device Manual Device Manual Device Manual Device Manual Device Manual
130* 138 127* 138 133* 139 74 72 70 70 78 79  ×  O
146* 123 133 122 132 120 79 68 73 68 72 70  ×   × 
136* 135 137* 131 134 129 102 106 103 104 100 103 O O
153 147 146 138 147* 134 57 52 51 50 58 50 O O
141 144 139 135 136* 133 82 82 79 82 80 79 O O
141 136 138* 133 130 128 93 93 92 92 92 91 O O
155 153 160 160 146 153 81 88 87 89 72* 90 O O

A total of 10 of 210 BP measurements that were discordant for high systolic blood pressure (SBP) or diastolic blood pressure (DBP) (> 135/85 mmHg) between the manual BP and the estimated BP from the ABPM device are marked with an asterisk (*). Hypertension was defined as ≥ 2 high SBP or DBP in three measurements.