Table 8.
Sensitivity, specificity, and agreement of combined Raybiotech+Healgen LFIA IgG with RT-PCR results.
| LFIA IgG combined | Total | |||
|---|---|---|---|---|
| + | - | |||
| RT-PCR | + | 54 | 19 | 73 |
| - | 0 | 122 | 122 | |
| Total | 54 | 141 | 195 | |
| Sensitivity | Specificity | Agreement | ||
| 74.0% | 100.0% | 90.3% | ||
Sensitivity, specificity, and agreement of combined Raybiotech+Healgen LFIA IgG with RT-PCR results.
| LFIA IgG combined | Total | |||
|---|---|---|---|---|
| + | - | |||
| RT-PCR | + | 54 | 19 | 73 |
| - | 0 | 122 | 122 | |
| Total | 54 | 141 | 195 | |
| Sensitivity | Specificity | Agreement | ||
| 74.0% | 100.0% | 90.3% | ||