Table 1.

UPLC®-MS procedures and conditions used for each platform

Platform 1 Platform 2 Platform 3
Column type UPLC BEH C18, 1.0 × 100 mm, 1.7 μm UPLC BEH C18, 2.1 × 100 mm, 1.7 μm UPLC BEH C18, 1.0 × 100 mm, 1.7 μm
Flow rate 0.14 ml/min 0.40 ml/min 0.14 ml/min
Solvent A H2O + 0.05% Formic Acid H2O + ACN + 10 mM Ammonium Formate 10 mM Ammonium Bicarbonate (pH= 8.8)
Solvent B ACN + 0.05% Formic Acid ACN+ Isopropanol + 10mM Ammonium Formate ACN
(%B), time 0%, 0 min 40%, 0 min 2%, 0 min
(%B), time 50%, 2 min 100%, 10 min 8%, 6.5 min
(%B), time 100%, 13 min 40%, 15 min 20%, 10 min
(%B), time 0%, 18 min 40%, 17 min 30%, 11 min
(%B), time - - 100%, 12 min
(%B), time - - 2%, 14 min
Column temperature 40 °C 60 °C 40 °C
Injection volume 2 μl 3 μl 2 μl
Source temperature 120 °C 120 °C 120 °C
Nebulisation N2 flow 600 l/hour 1000 l/hour 600 l/hour
Nebulisation N2 temperature 350 °C 500 °C 350 °C
Cone N2 flow 30 l/hour 30 l/hour 10 l/hour
Capillary voltage 2.8 kV 3.2 kV 3.2 kV
Cone voltage 50 V 30 V 30 V