. 2020 Sep 23;15(11):1622–1630. doi: 10.2215/CJN.07920520

Table 2.

Comparison of in-hospital adverse events between patients on dialysis who received subcutaneous implantable cardioverter defibrillators and those who received transvenous implantable cardioverter defibrillators

Adverse Event No. of Events per 100 Implants (95% CI) Odds Ratio (95% CI) P Value
Subcutaneous ICD Transvenous ICD Subcutaneous versus Transvenous ICD
Any adverse event 2.39 (1.62 to 3.51) 1.48 (1.02 to 2.14) 1.63 (0.95 to 2.82) 0.08
Death 0.82 (0.4 to 1.68) 0.31 (0.15 to 0.67) 2.62 (0.92 to 7.47)
Cardiac arrest 1.53 (0.9 to 2.61) 0.36 (0.17 to 0.76) 4.72 (1.71 to 11.17) 0.002
Hematoma 0.56 (0.3 to 1.05) 0.25 (0.11 to 0.57) 2.25 (0.80 to 6.32) 0.1
Hemothorax 0.07 (0.01 to 0.47) NA NA
Infection requiring antibiotics 0.10 (0.03 to 0.42) 0.03 (0.00 to 0.22) 3.29 (0.30 to 36.57) 0.3
Lead dislodgement 0.22 (0.08 to 0.59) 0.62 (0.34 to 1.11) 0.35 (0.11 to 1.13) 0.08
Tamponade 0.03 (0.00 to 0.22) NA NA
Pneumothorax 0.10 (0.03 to 0.42) NA NA
Stroke/TIA 0.08 (0.01 to 0.60) NA NA
Urgent cardiac surgery 0.07 (0.01 to 0.46) NA NA
AMI NA NA
Cardiac perforation NA NA

95% CI, 95% confidence interval; ICD, implantable cardioverter defibrillator; NA, not applicable; TIA, transient ischemic attack.