. 2020 Oct 13;8(10):1572. doi: 10.3390/microorganisms8101572

Table 1.

Sensitivities, specificities, and accuracies of eight severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) immunoglobulin G (IgG) or total antibody assays calculated from the results of two reference tests. The highest values are shown in green; the lowest values are marked in red.

(%) Sensitivity Sensitivity Specificity 3 Accuracy Accuracy
Reference test PCR 1 PRNT 2 PCR 1 PRNT 2
Assay\n 26 27 100 126 127
Abbott 96.2 96.3 99.0 98.4 98.4
Diasorin 84.6 88.9 96.0 93.7 94.5
Epitope 80.8 92.6 100.0 96.0 98.4
Euroimmun 80.8 88.9 97.0 93.7 95.3
Mikrogen 88.5 96.3 98.0 96.0 97.6
Roche 88.5 96.3 99.0 96.8 98.4
Viramed 88.5 96.3 99.0 96.8 98.4
Virion-Serion 96.2 96.3 100.0 99.2 99.2

1 SARS-CoV-2 RNA was previously detected in the PCR test. 2 SARS-CoV-2-neutralizing antibodies with a titer >1:10 were demonstrated in the plaque reduction neutralization test (PRNT). 3 100 stored sera from 2018/2019.