Skip to main content
. 2020 Nov 5;14:3801–3810. doi: 10.2147/OPTH.S274303

Table 3.

Mean Readings of Device 1 and Device 2, Difference and Mean Difference Between Device 1 and Device 2 and Limits of Agreement

Mean Device 1 Mean Device 2 Mean Difference Limit of Agreement
K1 42.8 43.02 −0.11 0.23 to −0.45
K2 43.99 44.21 −0.03 0.37 to −0.42
K1-back −6.08 −5.98 −0.11 −0.01 to 0.20
K2-back −6.41 −6.36 −0.05 0.07 to −0.18
K-max 44.62 45.49 −0.8 2.1 to −3.7
CCT 538.4 536 3.6 15.5 to −8.2
TLT 534.34 533 2.6 12.8 to −7.7
Q-front −0.37 −0.22 −0.16 0.03 to −0.35
Q-back −0.35 −0.78 −0.26 0.02 to −0.55
ACD 3.13 3.19 −0.05 0.04 to −0.14
ACA 39.65 42.44 −4.4 4.3 to −13.1
CD 11.72 12.2 −0.54 0.39 to −1.46
RMS-T 1.46 1.33 0.14 1.16 to −0.87
RMS-LOA 1.41 1.19 0.22 1.32 to −0.87
RMS-HOA 0.37 0.48 −0.11 0.11 to −0.33
SA 0.13 0.19 −0.05 0.19 to −0.29
Coma-0 0.03 −0.24 0.26 0.54 to −0.03
Coma-90 −0.12 0.003 −0.13 0.35 to −0.61
Trefoil-0 −0.04 0.014 −0.05 0.39 to −0.50
Trefoil-30 −0.08 −0.1 0.02 0.22 to −0.19

Abbreviations: SW, within-subject standard deviation; TRT, test–retest repeatability; CoV, coefficient of variation; K1, flat keratometry; K2, steep Keratometry; K Max, maximum keratometry; CCT, central corneal thickness; TLT, thinnest location thickness; Q-front, Q value on front surface at 6mm; Q-back, Q value on back surface at 6mm; ACD, anterior chamber depth; ACA, anterior chamber angle; CD, corneal diameter; RMS-T, total corneal aberration root mean square; RMS-LOA, root mean square of corneal low order aberrations; RMS-HOA, root mean square of corneal high order aberrations; SA, spherical aberrations; Coma-0, horizontal coma aberrations; Coma-90, vertical coma aberrations; Trefoil-0, oblique trefoil aberrations; Trefoil-30, vertical trefoil aberrations.