Table 1.

Overview of the technologies used for real-time motion monitoring.

Technology (section) Internal/external Dimensions Additional ionising radiation Tissue/tumour/surrogate Additional equipment to standard linac Online solution (vendor) if applicable
Infrared (2.1.1) External 1D No Patient surface No RPM (Varian) respiratory gating (figure 1(a))
6 DoF Fixation devices Yes IRLED (Brainlab)
Optical (2.1.2) External 6 DoF surface No Patient surface Yes Align RT (vision RT) (figure 1(b))/catalyst (C-RAD)
Spirometry (2.1.3) External 1D No Lung volume changes Yes ABC (Elekta) (figure 1(c))
Pressure belt (2.1.3) External 1D No Abdomen perimeter Yes Anzai (Anzai Medical) (figure 1(d)) Bellows
Thermistor (2.1.3) External 1D No Airflow temperature Yes Thermistor (non commercial)
kV/MV (2.2.2) Internal 3D triangulated Yes Markers (prostate) No MSKCC (non commercial)
kV/kV (2.2.2) Internal 3D triangulated Yes Markers (multi-site), vertebrae, Cranium Dedicated machine CyberKnife® (Accuray) (figure 2(c) top)
Markers (multi-site) Dedicated machine Vero (figure 2(c) bottom) (Brainlab and Mitsubishi, discontinued)
Markers (multi-site) Yes RTRT (non commercial)
Lung and liver tumours Yes Stereoscopic markerless monitoring (non commercial)
MV (2.2.3) Internal 2D beam’s eye view No Markers, lung tumour No No online solution
3D inferred Markers (prostate)
kV (2.2.3) Internal 3D inferred Yes Markers (multi-site), vertebrae, bronchi, lung tumours No KIM (non commercial, online only for prostate) and sequential stereoscopic (non commercial, online only for vertebrae)
6D inferred Markers KIM, not performed online
Hybrid (2.3) Internal with correlation model 3D Yes Markers (multi-site), lung tumours Dedicated machine CyberKnife® Synchrony (Accuray) (figure 2(c) top)
Markers (multi-site) Dedicated machine Vero (figure 2(c) bottom) (Brainlab and Mitsubishi, discontinued)
Markers (multi-site), cranium Yes ExacTrac (Brainlab) (figure 2(b))
Markers (lung) Yes RTRT  +  Anzai (non commercial)
Markers (liver) No COSMIK (non commercial)
Electromagnetic (2.4.1) Internal 3D No Markers (multi-site) Yes Calypso (Varian) and rayPilot (MicroPos Medical, only prostate) (figure 5)
Ultrasound (2.4.2) Internal 3D No Prostate, prostate bed Yes Clarity autoscan (Elekta) (figure 5(d))
Soft tissues Modified 4D ultrasound system (non commercial)
MR (2.5) Internal 2D cine (any orientation) No Tissues Dedicated machine Unity (Elekta), MRIdian (ViewRay) (figure 2(d))