Table 3.

Physical phantoms developed by research group.

Phantom (site) Deformable/anthropomorphic Motion reproducibility Main features
WashU (any site) (Malinowski et al 2007) No/no • Target accuracy (mean  ±  SD)  <  1 mm • 3D axis and independent 1D vertical axis
• Motorized platform to carry phantom or dosimetry equipment

LuCa (lung) (Perrin et al 2017) Yes (interior and exterior)/yes (high level of detail) • Stable end in/exhale (<1 mm) • Inflatable/deformable lungs, skeleton, muscles skin, solid heart, solid mobile tumour (can hold dosimetric films)
• Tumour position varied from day to day for a given intermediary in/exhale pressure • Motion actuated by an air pump inflating the lungs
• MR-compatible with visible deforming lung features

Lung (Cheung and Sawant 2015) Yes (interior and exterior)/yes (low level of internal detail) •  <2 mm day-to-day (ascribed to set-up) • Deformable external shell
• Latex foam insert for lungs
•  <  0.25 mm RMS intra-day • Rigid foam diaphragm actuated by the WashU motion stage

Lung (Biederer et al 2006, Remmert et al 2007) Yes (interior)/yes (animal heart and lungs, nodules, airways, no ribs) • Maximal diaphragm displacement precision (SD) 1.90 mm (on CT), 1.47 mm (on MR) • Porcine lung and heart explants with tracheal tube in saline solution, artificial pulmonary nodules
• Water-filled silicon diaphragm inflated or deflated by a water pump outside the MR room
• Reproducibility of intermediary phases not quantified • MR-compatible

Lung (Serban et al 2008) Yes (interior)/yes (only one lung with vasculature/airways features) • Within image resolution (0.7  ×  0.7  ×  1.25 mm3) • Lung (natural latex balloon filled with damp sponges) in water, thoracic cavity (Lucite), diaphragm (motor-actuated piston), tumour (Dermasol ellipsoid), vascular and bronchial bifurcation (nylon wires and Lucite beads)

Lung (Steidl et al 2012) Yes (exterior)/yes (low level of internal detail, cubic tumour) Target accuracy (mean  ±  SD)  =  0  ±  0.09 mm (input versus log files) • Artificial skeleton, rubber skin
• Tumour: PMMA cube with 20 slots for pinpoint ion chambers and 5 films
• Sternum-induced thoracic motion
• 6D robot-actuated tumour motion independent of thoracic motion

MAESTRO (lung) (Haas et al 2014) Yes (ribcage only)/yes (no vasculature) • Millimetre positioning precision • Mechanically actuated ribs, stationary lungs, trachea and spine in hermetic skin (to be filled with water)
• Inter-cycle reproducibility  <0.16 mm RMS • Robot-actuated tumour motion

ELPHA (liver) (Ehrbar et al 2019) Yes (interior)/yes (liver with vasculature) • Reproducibility  <0.32 mm RMS (inter- and intra-day) • Soft silicon liver with vasculature (can hold dosimetric devices)
• Static inferior plate and motor-driven superior plate
• Ultrasound and CT contrast