Imaging parameters and conditions for CBCT

10-Second CBCT 20-Second CBCT
Exposure parameters
    Tube voltage 70 kV 70 kV
    Pulse width 12.5 ms 12.5 ms
    Radiation exposure 1.20 μGy/pulse 1.20 μGy/pulse
    Binning/focus size/FOV 1 × 1/0.4 mm/22 cm
2 × 2/0.6 mm/42 cm
Acquisition parameters
    Angle step 0.8°/frame 0.4°/frame
    Angle 200° 200°
    Total frame 250 500
    Matrix size/binning 9602/2 × 2
10242/1 × 1
Injection condition
    Injection rate 1.0 mL/s 1.0 mL/s
    Concentration of contrast medium 14% 14%
    X-ray delay time 4.5 seconds 4.5 seconds
    Injection time of contrast medium 14.5 seconds 24.5 seconds