Abstract
Heart diseases and cardiovascular events are well-known side effects in left-sided breast irradiation. Deep inspiration breath hold (BH) combined with fast delivery techniques such as volumetric modulated arc therapy (VMAT) or tangential field-in-field (TFiF) can serve as a valuable solution to reduce the dose to the heart. This study aims to compare the impact of positioning errors in VMAT and TFiF plans for BH left-sided breast treatments. Fifteen left-sided breast patients treated in BH with TFiF technique were included in this retrospective study. For each patient, a second plan with VMAT technique was optimized. Eighteen setup variations were introduced in each of these VMAT and TFiF reference plans, shifting the isocenter along six different directions by 3, 5, and 10 mm. A total of 540 perturbed plans, 270 for each technique, were recalculated and analyzed. The dose distributions on the target and organs at risk obtained in the different perturbed scenarios were compared with the reference scenarios, using as dosimetric endpoints the dose-volume histograms (DVH). The results were compared using the Wilcoxon test. Comparable plan quality was obtained for the reference VMAT and TFiF plans, except for low doses to organs at risk for which higher values (p < 0.05) were obtained for VMAT plans. For TFiF plans, perturbations of the isocenter position of 3, 5, or 10 mm produced mean deviations of the target DVH dosimetric parameters up to −0.5, −1.0, and −5.2%, respectively; VMAT plans were more sensitive to positioning errors resulting in mean deviations up to −0.5, −4.9, and −13.9%, respectively, for the same magnitude of the above mentioned perturbations. For organs at risk, only perturbations along the left, posterior, and inferior directions resulted in dose increase with a maximum deviation of +2% in the DVH dosimetric parameters. A notable exception were low doses to the left lung and heart for 10 mm isocenter shifts for which the mean differences ranged between +2.7 and +4.1%. Objective information on how external stresses affect the dosimetry of the treatment is the first step towards personalized radiotherapy.
Keywords: left-sided breast, radiotherapy, volumetric modulated radiation therapy, breath hold, tangential field
Introduction
Breast cancer is the most common cancer in women worldwide. The standard of care includes conservative surgery or mastectomy as appropriate, followed by adjuvant radiotherapy. Radiotherapy is essential to improve local tumor control and overall survival; nevertheless, delivery of some dose to heart, lungs, and contralateral breast is unavoidable. Increased risk of fatal cardiac events, pneumonitis, or a second primary cancer of the breast has been largely reported (1–4). Given the occurrence of this disease even at a young age and the increased life expectancy, it is essential to limit as much as possible long-term complications.
Deep inspiration breath-hold (DIBH) in left-sided breast treatment increases the distance between the target and the heart as well as the part of lungs included in the treatment field. Many studies have been published on the benefits of DIBH showing how it enables minimized irradiation of nearby organs at risk while maintaining an adequate target dose coverage (5–8), and, therefore, has become the gold standard in clinical practice in many institutions (9–12).
Breast cancer treatment has historically been performed with two opposing non-divergent isocentric tangent fields, using six MV photon energies, with the addition of beam modifiers to homogenize the dose within the target (13, 14). The tangential field-in-field technique (TFiF), also referred to as forward intensity modulated radiation therapy, can be used as an alternative (15, 16); it consists of two open opposing tangential fields, but instead of wedges for target dose homogenization, additional fields (usually two to four) are manually created using a multileaf collimator (MLC). The main field and the subfields are therefore merged into one beam that includes several segments for the sequential irradiation. However, high intensity modulated radiotherapy techniques, such as volumetric modulated arc therapy (VMAT) and intensity modulated radiotherapy can achieve more homogeneous dose distribution within complex targets, such as concave-shaped breasts (17, 18). Moreover, inverse planning modalities have the advantage of optimizing the plan based on the clinical goals, offer fast planning options, and allow more tailored doses to organs at risk (OARs) (19, 20). Additionally, VMAT provides fast treatment delivery and is therefore an optimal candidate to pair with DIBH.
Others authors investigated the use of VMAT in the breath-hold treatment of left-sided breast radiotherapy and found unquestionable advantages for the heart dosimetry (21, 22). A paramount aspect to consider in these treatments is the uncertainty in inter-fraction patient re-positioning, which can lead to inaccuracies in the administered dose (23); when positioning errors occur, the steepness of the dose-effect curves can limit the effectiveness of highly modulated techniques such as VMAT, thus influencing the patient’s results both for local tumor control and for normal tissue complications. Moreover, daily shifts in patient setup are inevitable even with image guidance (24), and since it has become increasingly more evident in literature that imaging dose delivered to patients for pre-treatment image verification with cone-beam computed tomography (25), can be a concern (26), it is often not scheduled on a daily basis.
This work focuses on the use of BH with conformal TFiF and highly conformal VMAT irradiation techniques. Investigates whether, with an increased distance of the heart and lung from the target, VMAT and TFiF plans are robust against isocenter positioning errors; and it evaluates quantitatively their dosimetric impact on treatment plans for different magnitudes and directions.
Materials and Methods
This retrospective study included fifteen female early-stage pT1c-T2N1 left breast carcinoma patients receiving radiation therapy in DIBH between April and August 2019, after a breast-conserving surgery. The anatomical and clinical characteristics of the patients enrolled in this study are reported in Table 1 . The study was reviewed and approved by the Ethics Committee of our Hospital (approval number SCCHEC 02-2020003).
Table 1.
Patient # | Age* | Patient anatomy | Clinical characteristics | ||||
---|---|---|---|---|---|---|---|
PTV volume (cm3) | Heart volume (cm3) | Lung volume^ (cm3) | Body mass index (kg/m2) | Grade | pTNM | ||
1 | 47 | 441.5 | 360.2 | 3571.5 | 19.7 | G3 | T1cN0M0 |
2 | 23 | 88.3 | 408.2 | 4543.6 | 22.2 | G2 | T1cN0M0 |
3 | 31 | 509.8 | 422.4 | 2356.4 | 20.8 | G2 | T1cN0M0 |
4 | 49 | 383.0 | 637.5 | 3701.6 | 17.4 | G1 | T1cN0M0 |
5 | 49 | 486.7 | 429.5 | 4150.9 | 18.8 | G1 | T2N0M0 |
6 | 34 | 598.6 | 469.9 | 2399.9 | 21.9 | G3 | T1cN0M0 |
7 | 39 | 500.8 | 518.8 | 3963.7 | 21.8 | G2 | T1cN0M0 |
8 | 31 | 638.3 | 540.9 | 4022.7 | 20.8 | G2 | T1cN0M0 |
9 | 56 | 389.9 | 595.9 | 3146.3 | 25.0 | G1 | T2N0M0 |
10 | 33 | 247.5 | 447.1 | 3112.0 | 19.6 | G1 | T1cN0M0 |
11 | 48 | 484.7 | 459.2 | 3627.1 | 20.5 | G2 | T1cN0M0 |
12 | 44 | 654.4 | 342.3 | 4473.2 | 21.0 | G2 | T1cN0M0 |
13 | 55 | 497.8 | 577.2 | 3625.5 | 19.8 | G2 | T1cN0M0 |
14 | 33 | 776.2 | 544.5 | 4236.4 | 21.3 | G3 | T1cN1M0 |
15 | 58 | 461.0 | 575.6 | 3492.8 | 19.2 | G1 | T1aN0M0 |
Mean | 42 | 510.5 | 488.6 | 3628.2 | 20.7 | – | – |
Median | 44 | 497.8 | 469.9 | 3627.1 | 20.8 | – | – |
PTV, planning target volume; pTNM, pathological tumor node metastasis stage; *at the time of radiotherapy; ^volumes at breath hold setup.
Clinical Workflow
In DIBH treatment, patients must be able to inspire and then maintain inspiration during each treatment. At the planning computed tomography (CT), time was dedicated to patient training. Patients were immobilized in a supine position with a standard commercial breast board (WingSTEP, IT-V, Innsbruck, Austria) that allowed them to comfortably rest their arms behind their heads; an anterior and two lateral tattoos were performed at the free breathing setup position corresponding to the laser cross. Patients were then asked to inhale so as to swell the upper chest and to hold their breath, and three additional skin marks were performed at the laser cross to set the breathing retention setup. To be eligible for DIBH treatment, patients must be able to hold their breath for at least 25 s and to replicate the breath retention setting five times in succession. The setup reproducibility at the CT training was verified by checking the alignment of tattoos/lasers in the free breathing setup, followed by the alignment of marks/lasers in breath hold; moreover, the height of the breath-hold lateral tattoos above the couch top during breath-hold were registered and compared with the height of the lateral tattoos on the CT scan to confirm that a consistent breath-hold was performed. CT scans with 3 mm slice thickness were acquired in breath-hold with a 16-slice Brilliance Big Bore CT (Philips Medical Systems, Cleveland, OH). In the treatment room, the patient is first aligned in free breathing; then, in the same way as at the CT training session, during the breath-hold retention the accuracy and repeatability of the breath-hold setup is verified and, for each treatment field, the field border is marked on the patient skin. The patient was asked to perform a breath-hold; using treatment room cameras, therapists at the LINAC’s console administered the radiation when the light field and previously marked field borders were correlated. Electronic portal images acquired for each beam in breath-hold before delivery were matched online to digitally reconstructed radiographs. The voluntary breath-hold technique described above has been implemented and used in the clinical practice by the Royal Marsden Hospital in London (UK) (27). Barlett et al. (28) found comparable results in the reproducibility and normal tissue sparing of this technique and that of ABC (active breathing coordinator); however, they noted that patients found the voluntary DIBH technique more comfortable and less claustrophobic.
Experienced radiation oncologists from the breast oncology department outlined the target and the OARs on the CT images dataset imported into Pinnacle 3TM Version 9.10 (Philips Medical Systems, Eindhoven, the Netherlands). The clinical target volume (CTV) consisted of the whole left breast, excluding pectoralis muscles, chest wall muscles, and ribs. The planning target volume (PTV) was an isotropic expansion of the CTV with a 3 mm margin in all directions; the first 5 mm inside the body external contour were excluded both from the CTV and from the PTV. OAR delineations were performed according to the breast cancer atlas for the radiation therapy planning consensus definitions (29). An hypo-fractionated regimen as standard of care for early stage breast cancer (30) has been adopted in our center with a prescribed dose (Dp) of 42.56 Gy delivered in 16 fractions (31, 32) over 3 weeks. The plan has been optimized to achieve minimum 95% of the PTV covered by 95% isodose line and a mean dose to PTV equal to the Dp; hotspots should not exceed 107%, although they were considered acceptable if 2 cm3 of the target received 110% of the Dp. For OARs standard dose limits were used (33, 34) aiming at keeping a heart Dm under 2 Gy and less than 15% of the left lung receiving less than 20 Gy (10).
Tangential Field in Field and Volumetric Modulated Arc Therapy Reference Plans
An Elekta Infinity (Elekta, Stockholm, Sweden) LINAC mounting a 5 mm MLC was used for the treatments. The TFiF treatment plan delivered to the patients included in this study was performed with a 6 MV beam and consist of two opposing tangential fields with gantry angle between 300° and 315° for the medial beam and 120° and 135° for the lateral beam, each including two or three sub-segments. The main segment that corresponds to the whole tangential field consisted of about 80% of the monitor units (MU). TFiF is a forward plan and its clinical implementation is closely related to conventional planning. Manual definition of the segments leads to intuitive choices for the segment shapes on the beam’s-eye-view option of the planning system. Two opposite open tangential fields were initially created, equal weights were assigned and the corresponding dose distribution was calculated. High-spot volumes were created from the isodoses, and subfield of the tangential beams were added manually conforming the leaves of the MLC to cover the hot-spot volumes; the dose distribution was then recalculated and the weights of field and subfields were adjusted to improve the dose homogeneity. This process was repeated until the accurate dose distribution was reached. The number of subfields varied between two and three. The plan was constrained to a delivery time for each beam shorter than 20 s to introduce a safety margin with respect to the inclusion criteria. The plan was calculated using the full collapsed cone convolution algorithm and a grid calculation size of 3 mm. The TFiF plan clinically delivered, represents the TFiF reference plan (TFiFref) for the purpose of this research.
For each case, a second plan was retrospectively implemented for the purpose of our research, using the VMAT technique and the same LINAC but with the 6 MV flat flattening free beam energy and dose rate of 1,400 MU/min. This choice was made to guarantee the shortest delivery time, which is advantageous for a patient treated in breath hold, with a target metric comparable to the one obtainable with flattened filter beams (35–37); we will refer to it as VMAT reference plan (VMATref). The plan consisted of two 40° partial arcs with 300°/340° and 100°/140° as start/stop angles, respectively, with a variability of 10° in the start/stop gantry angles based on the specifics of the patient’s treatment plan; the plan was optimized following the same criteria for the maximum delivery time, OARs dose constraints, target dose coverage. The same radiation oncologist approved the TFiFref and, successively, the VMATref plans.
Plans Perturbations Comparisons
Eighteen setup variations were introduced on each VMATref and TFiFref plan, shifting the isocenter from its reference position in the superior (S), inferior (I), left (L), right (R), anterior (A), and posterior (P) directions with respect to the patient couch view from the feet of the patient by 3, 5, and 10 mm. A total of 540 treatment plans were recalculated with these simulated positioning errors on the planning CT without changing any parameter other than the position of the isocenter. Dose volume histogram (DVH) endpoints were used to compare the impact of the variations in setup on the dosimetry of the VMAT and TFiF plans; in particular, D95, D98, D2cc, and Dm for the CTV; V5, V10, V25, and Dm for the heart; V5, V20, and Dm for the left lung; and V5 and Dm for the right breast, were used, where Dx represented the dose (in Gy) received by x% of the volume, Vy the volume (in percentage) receiving y Gy, and D2cc the dose in Gy received by a volume of 2 cm2. Absolute differences ΔDx, ΔVy, and ΔDm were calculated by subtracting the reference value Dx, Vy, Dm, respectively, from the corresponding perturbed value.
Datasets were compared using the Wilcoxon signed-rank test. A p value of less than 0.05 was considered statistically significant; IBM SPSS v20 software (IBM, Armonk, US) was used for the analysis.
Results
CTV and OARs DVH dosimetric parameters (mean and range values), obtained for the non-perturbed reference VMAT and TFiF plans, are shown in Table 2 . Target coverage was clinically acceptable for both techniques and for all the cases studied, with a minimum D95 value of 40.4 Gy, Dm equal to Dp ± 0.3 Gy, and a maximum D2cc value of 45.4 Gy. Similarly, standard dose constraints were satisfied for all OARs; heart Dm resulted <2 Gy, with the exception of one case that achieved a Dm of 3.2 Gy; and left lung V20 < 15% with a maximum value registered of V20 = 16.4%. Dm for the right breast was <2.4 Gy with a mean value of 1.5 Gy.
Table 2.
VMAT | TFiF | p value | ||
---|---|---|---|---|
CTV | D95 (Gy) | 41.0 (40.4–41.4) | 40.9 (40.4–41.4) | 0.535 |
D98 (Gy) | 40.3 (39.9–40.7) | 40.3 (39.6–40.4) | 0.430 | |
D2cc (Gy) | 43.9 (43.3–45.5) | 44.4 (43.4–45.4) | 0.327 | |
Dm (Gy) | 42.6 (42.3–42.8) | 42.8 (42.5–42.9) | 0.342 | |
Heart | V5 (%) | 4.2 (3.8–6.0) | 3.1 (1.1–5.2) | 0.052 |
V10 (%) | 1.2 (0.6–2.6) | 1.3 (0.4–2.5) | 0.061 | |
V25 (%) | 0.2 (0.0–0.4) | 0.3 (0.0–0.8) | 0.370 | |
Dm (Gy) | 1.6 (0.5–3.2) | 1.5 (0.5–2.9) | 0.350 | |
Left lung | V5 (%) | 29.0 (23.2–34.5) | 25.7 (21.0–30.0) | 0.037 |
V10 (%) | 20.0 (15,3–24.8) | 19.2 (15.3–22.3) | 0.132 | |
V20 (%) | 12.4 (9.1–16.4) | 12.8 (9.2–16.4) | 0.230 | |
Dm (Gy) | 6.6 (5.3–8.2) | 6.6 (5.3–7.9) | 0.123 | |
Right breast | V5 (%) | 1.7 (0.2–3.0) | 1.1 (0.0–2.2) | 0.034 |
Dm (Gy) | 1.5 (0.8–2.4) | 1.5 (0.5–2.4) | 0.321 |
DVH, dose volume histogram; VMAT, volumetric modulated arc therapy; TFiF, tangential field in field; CTV, clinical target volume; Dx, dose in Gy received by x% of the volume; Dm, mean dose; Vx, volume receiving x Gy; D2cc, dose in Gy received by a volume of 2 cube centimeters.
The comparison of the target coverage obtained with the reference vs. the perturbed plans is shown in Table 3 ; the mean and range of the absolute differences of CTV DVH dosimetric parameters obtained are reported for different magnitudes and directions of the isocenter shifts.
Table 3.
Isocenter shift Dir; mm | CTV, ΔD95 (Gy) | CTV, ΔD98 (Gy) | CTV, ΔDm (Gy) | |||
---|---|---|---|---|---|---|
TFiF | VMAT | TFiF | VMAT | TFiF | VMAT | |
I; 3 | −0.1 (−0.1; −0.3) | −0.1 (0.0; −0.1) | −0.1 (0.0; −0.3) | −0.1 (0.0; −0.2) | 0.0 (0.0; −0.1) | −0.1 (0.0; −0.3) |
S; 3 | −0.2 (−0.1; −0.4) | −0.2 (−0.1; −0.3) | − 0.2 (−0.1; −0.4) | −0.2 (−0.1; −0.4) | −0.2 (−0.1; −0.5) | −0.2 (−0.1; −0.4) |
L; 3 | −0.2 (−0.1; −0.4) | −0.3 (−0.3; −0.4) | −0.2 (−0.1; −0.3) | −0.3 (−0.2; −0.5) | −0.2 (−0.1; −0.4) | −0.1 (0.0; −0.3) |
R; 3 | −0.2 (0.0; −0.3) | −0.2 (−0.1; −0.4) | −0.2 (0.0; −0.3) | −0.3 (−0.1; −0.6) | −0.2 (0.0; −0.3) | −0.2 (−0.1; −0.3) |
P; 3 | −0.1 (−0.1; −0.2) | −0.1 (0.0; −0.2) | −0.1 (0.0; −0.1) | −0.1 (−0.1; −0.2) | −0.1 (0.0; −0.2) | −0.2 (−0.2; −0.3) |
A; 3 | −0.3 (−0.2; −0.5) | −0.4 (−0.2; −0.6) | −0.4 (−0.2; −0.6) | −0.4 (−0.3; −0.5) | −0.3 (−0.1; −0.4) | −0.3 (−0.1; −0.3) |
All Dir | −0.2 (−0.0; −0.5) | −0.2 (−0.0; −0.6) | −0.2 (0.0: −0.6) | −0.2 (0.0: −0.6) | −0.2 (0.0; −0.5) | −0.2 (0.0; −0.4) |
I; 5 | −0.2 (−0.1; −0.3) | −0.8 (−0.4; −1.5) | −0.2 (0.0; −0.2) | −1.0 (−0.6; −1.4) | −0.1 (0.0; −0.3) | −0.3 (−0.1; −0.3) |
S; 5 | −0.3 (−0.1; −0.4) | −1.7 (−1.1; −2.7) | −0.4 (−0.2; −0.6) | −2.2 (−2.0; −2.7) | −0.2 (−0.1; −0.4) | −0.2 (−0.1; −0.4) |
L; 5 | −0.3 (−0.2; −0.4) | −0.9 (−0.5; −1.9) | −0.3 (−0.2; −0.4) | −1.1 (−0.7; −1.5) | −0.2 (−0.1; −0.4) | −0.3 (−0.1; −0.4) |
R; 5 | −0.4 (−0.2; −0.5) | −2.1 (−1.9; −2.7) | −0.4 (−0.2; −0.6) | −2.9 (−2.0; −3.6) | −0.2 (−0.1; −0.3) | −0.2 (0.0; −0.3) |
P; 5 | −0.1 (0.0; −0.3) | −1.0 (−0.8; −1.6) | −0.2 (−0.2; −0.3) | −1.6 (−1.2; −1.8) | −0.1 (0.0; −0.2) | −0.1 (0.0; −0.3) |
A; 5 | −0.5 (−0.4; −0.8) | −2.6 (−2.2; −3.0) | −0.6 (−0.4; −1.4) | −3.2 (−2.4; −4.5) | −0.3 (−0.2; −0.5) | −0.3 (−0.2; −0.5) |
All dir | −0.3 (−0.1; −0.8) | −1.5 (−0.5; −2.9) | −0.4 (0.0: −1.4) | −2.0 (−0.6; 4.5) | −0.2 (0.0; −0.5) | −0.2 (−0.1; −0.5) |
I; 10 | −0.5 (−0.1; −0.6) | −1.4 (−1.2; −1.7) | −1.1 (−0.7; −1.4) | −2.7 (−2.2; −3.4) | −0.3 (−0.3; −0.4) | −0.9 (−0.4; −1.1) |
S; 10 | −1.1 (−0.9; −1.3) | −3.4 (−3.0; −3.6) | −2.0 (−1.3; −3.0) | −5.3 (−4.5; −6.1) | −0.5 (−0.4; −0.9) | −0.8 (−0.4; −1.1) |
L; 10 | −0.7 (−0.6; −1.0) | −2.1 (−1.6; −2.4) | −0.9 (−0.7; −1.3) | −3.2 (−2.1; −4.3) | −0.4 (−0.2; −0.6) | −0.9 (−0.9; −1.3) |
R; 10 | −2.2 (−1.9; −2.6) | −6.3 (−5.7; −6.3) | −3.2 (−2.8; −3.8) | −8.9 (−6.9; −10.1) | −0.6 (−0.4; −0.8) | −1.3 (−0.8; −1.1) |
P; 10 | −0.7 (−0.4; −1.6) | −2.0 (−1.6; −2.2) | −1.2 (−0.6; −1.8) | −3.4 (−2.9; −3.9) | −0.2 (−0.1; −0.3) | −0.9 (−0.7; −1.0) |
A; 10 | −2.6 (−2.0; −3.3) | −7.3 (−5.3; −8.1) | −3.8 (−3.5; −4.3) | −10.1 (−6.5; −12.1) | −0.8 (−0.5; −1.1) | −1.2 (−0.7; −1.4) |
All dir | −1.3 (−0.1; −3.3) | −3.8 (−1.2; −8.1) | −2.1 (−0.6; −4.3) | −5.6 (−2.1; −12.1) | −0.5 (−0.1; −1.1) | −1.0 (−0.4; −1.4) |
VMAT, volumetric modulated arc therapy; TFiF, tangential field in field; CTV, clinical target volume; Dx, dose in Gy received by x% of the volume; Dm, mean dose; Vx, volume receiving x Gy. I, inferior; S, superior; P, posterior; L, left; R, right; A, anterior; P. posterior; Dir, direction.
In gray data that with mean absolute difference ≤2%, in bold > 3%.
For a 3 mm perturbation, both techniques showed a mean absolute difference ΔD95 and ΔD98 in all directions ≤−0.2 Gy (0.5%), with maximum value of −0.4 Gy (1.0%) in the A direction. Isocenter shifts of 5 mm for TFiF perturbed plans produced mean absolute difference ΔD95 and ΔD98 of −0.3 Gy (−0.7%), and −0.4 Gy (−1.0%), respectively, with maximum differences of −0.5 Gy (−1.2%) and −0.6 Gy (−1.5%), respectively, in the A direction; for VMAT perturbed plans, larger mean ΔD95 and ΔD98 values were registered: −1.5 Gy (−3.7%), and −2.0 Gy (−5.0%), respectively, with maximum differences in the A direction of −2.6 Gy (−6.3%) and 3.2 Gy (7.9%), respectively. For 10 mm isocenter shifts, mean ΔD95 and ΔD98 of −1.3 Gy (−3.2%) and −2.1 Gy (−5.2%) respectively, were registered for TFiF perturbed plans, whereas VMAT perturbed plans produced larger values of −3.8 Gy (−9.3%) and −5.6 Gy (−13.9%), for ΔD95 and ΔD98, respectively. A and R directions contributed most to worsening the target dosimetry with ΔD95 of −2.6 Gy (−6.4%), and −2.2 Gy (−5.4%), respectively, for TFiF plans, and −7.3 Gy (17.8%) and −6.3 Gy (−15.4%) for VMAT plans; similarly, for ΔD98 differences of −3.8 Gy (9.4%), and −3.2 Gy (7.9%), in the A and R direction, respectively, were obtained for TFiF plans, whereas −10.1 Gy (25.1%), and −8.9 Gy (22.1%), respectively, were reported for VMAT plans.
Only isocenter shifts in the L, P, and I directions increased the dose to OARs. Shifts in the other three directions (R, A, S) had the effect of moving the OARs further away from the treatment field, thus decreasing the dose received. In Figure 1 the heart ΔV5 and ΔV25 and in Figure 2 the lung ΔV5 and ΔV20 are plotted for each isocenter shift direction and magnitude. The mean and range values of the absolute differences for the DVH parameters between the reference plans and the perturbed plans along the L, P, and I directions are reported in Table 4 for the heart, the left lung, and the right breast. The mean absolute difference across all OARs, treatment techniques, and perturbation values remain <2%, with the exception of the heart and left lung ΔV5, which rises for isocenter shifts of 10 mm to 3.9 and 2.7%, respectively, for TFiF, and 4.1 and 3.0%, respectively, for VMAT plans.
Table 4.
10 mm | 5 mm | 3 mm | |||||||
---|---|---|---|---|---|---|---|---|---|
VMAT | TFiF | VMAT | TFiF | VMAT | TFiF | ||||
Heart | ΔV5 (%) | 4.1 (2.9–5.5) | 3.9 (3.0–4.7) | 1.5 (2.0–4.1) | 1.4 (1.1–2.0) | 0.1 (0.0–0.2) | 0.0 (0.0–0.1) | ||
ΔV10 (%) | 1.9 (1.7–34.4) | 1.9 (1.5–3.3) | 1.3 (1.0–2.0) | 1.3 (1.0–1.8) | 0.1 (0.0–0.3) | 0.1 (0.0–0.3) | |||
ΔV25 (%) | 1.8 (1.6–2.5) | 1.5 (1.3–2.3) | 0.9 (0.7–1.4) | 0.7 (0.4–1.5) | 0.0 (0.0–0.2) | 0.1 (0.1–0.2) | |||
ΔDm (%) | 0.5 (0.7–1.3) | 0.5 (0.8–1.4) | 0.4 (0.4–0.6) | 0.2 (0.1–0.4) | 0.1 (0.0–0.3) | 0.0 (0.0–0.1) | |||
Left lung | ΔV5 (%) | 3.0 (2.4–5.6) | 2.7 (2.3–4.6) | 1.6 (1.2–2.5) | 1.3 (1.0–2.6) | 0.2 (0.2–0.6) | 0.2 (0.1–0.3) | ||
ΔV20 (%) | 1.9 (1.7–4.0) | 1.8 (1.4–3.5) | 1.2 (1.0–2.3) | 1.0 (0.7–1.9) | 0.2 (0.1–0.4) | 0.1 (0.2–0.3) | |||
ΔDm (%) | 0.8 (0.6–1.4) | 0.8 (0.5–1.4) | 0.5 (0.4–1.0) | 0.5 (0.3–0.9) | 0.4 (0.3–0.5) | 0.4 (0.3–0.4) | |||
Right breast | ΔV5 (%) | 0.1 (0.0–0.1) | 0.1 (0.0–0.1) | 0.0 (0.0–0.0) | 0.0 (0.0–0.1) | 0.0 (0.0–0.0) | 0.0 (0.0–0.1) | ||
ΔDm (Gy) | 0.1 (0.0–0.2) | 0.1 (0.0–0.1) | 0.0 (0.0–0.1) | 0.0 (0.0–0.1) | 0.0 (0.0–0.1) | 0.0 (0.0–0.1) |
VMAT, volumetric modulated arc therapy; TFiF, tangential field in field; Dx, dose in Gy received by x% of the volume; Dm, mean dose; Vx, volume receiving x Gy.
In gray data with mean difference ≤ 2%, in bold >3%.
Discussion
There is now sufficient evidence that DIBH in left-sided breast radiotherapy allows for dosimetric sparing of OARs (5–8). Historically, treatment plans of the breast were performed with opposing tangent fields, subsequently optimized with the field-in-field technique. In our institution, DIBH TFiF is offered to every left-sided breast patient undergoing a radiotherapy treatment after conservative surgery. The availability of a new LINAC with high dose rate drove us to consider using VMAT for these treatments and to compare them with our standard of care (TFiF). Furthermore, the idea was also to compare the response of these two irradiation techniques when perturbed by incorrect isocenter positioning, verifying whether using them in BH could have a minor effect on the dosimetric impact on the organs at risk. Among the patients studied, the BMIs were almost all within the standard level value, correctly representing the female population afferent to our hospital; in any case, BMI, heart, and total lung volumes have been reported in the literature as having a minimal impact on the target dose and lung dosimetry (38).
Both the reference TFiF and the VMAT plans provided adequate and similar CTV dose coverage (p > 0.05) and OAR sparing (p > 0.05), except for the low doses (V5) for which higher values were obtained in VMAT plans (p > 0.05). The low-dose bath exposure of healthy structures is a well-known limitation of VMAT in breast cancer treatment (39); nevertheless, the difference was small, the values obtained were well above the OAR constraints, and the VMAT plans were considered clinically acceptable. These results are in agreement with previously published studies (40).
When perturbations were introduced, however, TFiF techniques guaranteed an accurate target coverage for isocenter shifts up to 5 mm with deviations of the target DVH dosimetric parameters <1.0%, whereas VMAT plans seemed more sensitive to positioning errors registering mean deviations of −3.7% and −5.0% for D95 and D98, respectively. In Table 3 it is possible to observe that each isocenter perturbation has a dosimetric impact on the CTV, which is larger for VMAT plans than for TFiF plans, and this is amplified for isocenter’ perturbations of 10 mm, for which mean ΔD95 of −9.3 vs. −3.2% respectively, and mean ΔD98 of −13.9 vs. −5.2%, respectively, were registered. Moreover, perturbations in the A and R directions most affected the target dosimetry for both techniques. Nevertheless, for VMAT plans perturbed with isocenter shifts of 10 mm, all the directions have a significant impact on the plan dosimetry, contributing to D95 deviations between −3.4% (in the I direction), and −17.8% (in the A direction); larger deviations, between −6.7, and −25.1%, in the I and A directions, respectively) were registered for D98. These high deviations on VMAT plans target dosimetry reported for each direction of the perturbation make a customized solution difficult. For TFiF plans, instead, mean ΔD95 and ΔD98 are smaller, and only the A and R directions, and the A, R, and I directions, respectively, contribute with deviations larger than 3%. This suggests it may be possible to mitigate the dosimetric deviations to below 3% in those specific directions for isocenter’s positioning errors of 10 mm.
For the OARs, only the L, P, and I isocenter shift directions contributed to increase the dose received, bringing the treatment field closer to the OARs. Considering only these directions that worsen the OARs dosimetry, we obtained dose values with deviations of less than 2% from the reference plan doses, with the exception of 10 mm perturbation, for which mean differences between 3.7 and 4.1% were registered for the low doses. Left sided-breast BH irradiation has therefore the advantage not only of limiting the dose to the heart and lung if compared with free breathing delivery, but also of being more robust against possible heart and lung overdosage in case of unexpected isocenter misplacement. The benefit of post-operative RT for breast cancer patients, in term of reduced risk of recurrence, has been demonstrated (41, 42); nevertheless it is well known that concurrent heart irradiation leads to an increased risk of heart disease (43), with evidence of increased risk of death of 3% per Gy of the Dm (44). Recent literature (45) reports no correlations between incidental heart irradiation and cardiac mortality, nevertheless there is no consensus yet on cardiac irradiation induced mortality in breast radiotherapy. Moreover, the well-known second cancer risk for contralateral breast and lung (46) force us to monitor the treatment delivery to ensure that the OAR delivered dose corresponds to the planned ones. One possible application would be in systems such as surface imaging, often used to detect setup errors, which offer real-time monitoring and beam delivery interruption if patient’s positions exceed their tolerance limits (11). Personalized thresholds in the different directions based on the results presented here would ensure accurate dose delivery.
Among this study’s limitations, it is important to highlight that a single institution was involved and the results were obtained according to our center’s working protocols. The study is focused on early breast cancer patients performing radiotherapy without an implanted tissue expander, further investigations are needed for this case. Contouring of the regions of interest and approval of treatment plans were performed by a single experienced radiation oncologist for consistency in comparing the different treatment modalities.
Jensen et al. (47) analyzed the influence of localization errors on VMAT and 3DCRT breast plans using weekly offline imaging and throughout the treatment session; their findings, representative of the workflow of their Center, show that perturbed dose calculated on the treatment data were less variable for VMAT than for 3DCRT plans. Personalized radiotherapy means being able to adapt the clinical workflow to the way the individual institute works. The knowledge and therefore the study of how the system responds to external stress is the first step towards this ambitious goal.
Left sided breast treatment performed in breath hold maintains the heart and lung doses close to the planned values (deviations < 2%) for setup errors up to 10 mm from the isocenter and for both delivery techniques, except for heart and lung low doses that increase up to 4.1%. TFiF technique guarantees an accurate target coverage for isocenter shifts of 3 mm, and 5 mm with dose deviations < 1%; for isocenter shifts of 10 mm, target dosimetric parameter deviations ranged between −3.2 and −5.2%; the main contribution to these deviations is due to perturbations in specific directions, leaving open the possibility to limit them with appropriate personalized management of the treatment. VMAT plans seem more sensitive to positioning errors, showing mean target dosimetric parameters deviations up to 5% for 5 mm isocenter shifts, and up to 13.9% for 10 mm isocenter shifts, leaving less room to control the target dosimetry. The evaluation of the dosimetric impact when the ideal system is perturbed therefore remains of primary importance. Quantitative information may support radiation oncologists in setting up a personalized radiotherapy.
Data Availability Statement
The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.
Ethics Statement
The studies involving human participants were reviewed and approved by Ethics Committee of Sichuan Cancer Hospital (approval number SCCHEC 02-2020003). The patients/participants provided their written informed consent to participate in this study.
Author Contributions
Design of the research: LCO, YZ, JL. Treatment plans: DZ, JW, XX, PD, ML. Statistical Analysis: DF, JL. Manuscript preparation: DF, XX, JW. Manuscript writing: LCO, YZ. Manuscript final revision: DF, DP, JL. All authors contributed to the article and approved the submitted version.
Funding
This research was supported by Chengdu Science and Technology Project (2019-YF09-00095-SN).
Conflict of Interest
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
The handling editor declared a past co-authorship with one of the authors DF.
Abbreviations
DIBH, deep inspiration breath hold; TFiF, tangential field in field; MLC, multi leaf collimator; VMAT, volumetric modulated arc therapy; OARs, organs at risk; CT, computed tomography; CTV, clinical target volume; PTV, planning target volume; Dm, mean dose; DVH, dose volume histogram.
References
- 1. Boero IJ, Paravati AJ, Triplett DP, Hwang L, Murphy JD. Modern Radiation Therapy and Cardiac Outcomes in Breast Cancer. Int J Radiat Oncol Biol Phys (2016) 94(4):700–8. 10.1016/j.ijrobp.2015.12.018 [DOI] [PubMed] [Google Scholar]
- 2. Darby SC, Ewertz M, Mcgale P, Bennet AM, Blomgoldman U, Brønnum D, et al. Risk of ischemic heart disease in women after radiotherapy for breast cancer. N Engl J Med (2013) 368(11):987–98. 10.1056/NEJMoa1209825 [DOI] [PubMed] [Google Scholar]
- 3. Cuzick J, Stewart H, Rutqvist LE, Houghton J, Host H. Cause-specific mortality in long-term survivors of breast cancer who participated in trials of radiotherapy. J Clin Oncol (1994) 12(3):447–53. 10.1200/JCO.1994.12.3.447 [DOI] [PubMed] [Google Scholar]
- 4. Stovall M, Smith SA, Langholz BM, Boice JD, Shore RE, Andersson M, et al. Dose to the Contralateral Breast From Radiotherapy and Risk of Second Primary Breast Cancer in the WECARE Study. Int J Radiat Oncol Biol Phys (2008) 17(72(4):1021–30. 10.1016/j.ijrobp.2008.02.040 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Swanson T, Grills IS, Ye H, Entwistle A, Teahan M, Letts N, et al. Six-year Experience Routinely Using Moderate Deep Inspiration Breath-hold for the Reduction of Cardiac Dose in Left-sided Breast Irradiation for Patients With Early-stage or Locally Advanced Breast Cancer. Am J Clin Oncol (2013) 36(1):24–30. 10.1097/COC.0b013e31823fe481 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Latty D, Stuart KE, Wang W, Ahern V. Review of deep inspiration breath-hold techniques for the treatment of breast cancer. J Med Radiat Sci (2015) 62(1):74–81. 10.1002/jmrs.96 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. Remouchamps VM, Vicini FA, Sharpe MB, Kestin LL, Martinez AA, Wong JW. Significant reductions in heart and lung doses using deep inspiration breath hold with active breathing control and intensity-modulated radiation therapy for patients treated with locoregional breast irradiation. Int J Radiat Oncol Biol Phys (2003) 55(2):392–406. 10.1016/S0360-3016(02)04143-3 [DOI] [PubMed] [Google Scholar]
- 8. Pham TT, Ward R, Latty D, Owen C, Gebski V, Chojnowski J, et al. Left-sided breast cancer loco-regional radiotherapy with deep inspiration breath-hold: Does volumetric-modulated arc radiotherapy reduce heart dose further compared with tangential intensity-modulated radiotherapy? J Med Imaging Radiat Oncol (2016) 60(4):545–53. 10.1111/1754-9485.12459 [DOI] [PubMed] [Google Scholar]
- 9. Mika J, Laaksomaa M, Jani P, Mikko H, Helmi L, Turkka L, et al. Effect of multiple breath hold reproducibility on treatment localization and dosimetric accuracy in radiotherapy of left-sided breast cancer with voluntary deep inspiration breath hold technique. Med Dosim (2017) 42:177–84. 10.1016/j.meddos.2017.02.004 [DOI] [PubMed] [Google Scholar]
- 10. Vikström J, Hjelstuen MHB, Mjaaland I, Dybvik KI. Cardiac and pulmonary dose reduction for tangentially irradiated breast cancer, utilizing deep inspiration breath-hold with audio-visual guidance, without compromising target coverage. Acta Oncol (2011) 50(1):42–50. 10.3109/0284186X.2010.512923 [DOI] [PubMed] [Google Scholar]
- 11. Xiao A, Crosby J, Malin M, Malin H, Kang H, Washington M, et al. Single-institution report of setup margins of voluntary deep-inspiration breath-hold (DIBH) whole breast radiotherapy implemented with real-time surface imaging. J Appl Clin Med Phys (2018) 19(4):205–13. 10.1002/acm2.12368 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Nissen HD, Appelt AL. Improved heart, lung and target dose with deep inspiration breath hold in a large clinical series of breast cancer patients. Radiother Oncol (2013) 106(1):28–32. 10.1016/j.radonc.2012.10.016 [DOI] [PubMed] [Google Scholar]
- 13. Dobbs J, Barrett A, Ash D. Breast. In: Koster J, editor. Practical Radiotherapy Planning. Great Britain: Arnold; (1999). p. 200–13. [Google Scholar]
- 14. Krasin M, Mccall A, King S, Olson M, Emami B. Evaluation of a standard breast tangent technique: a dose-volume analysis of tangential irradiation using three-dimensional tools. Int J Radiat Oncol Biol Phys (2000) 47(2):327–33. 10.1016/S0360-3016(00)00449-1 [DOI] [PubMed] [Google Scholar]
- 15. Onal C, Sonmez A, Arslan G, Oymak E, Kotek A, Efe E, et al. Dosimetric comparison of the field-in-field technique and tangential wedged beams for breast irradiation. Jpn J Radiol (2012) 30(3):218–26. 10.1007/s11604-011-0034-7 [DOI] [PubMed] [Google Scholar]
- 16. Hidekazu T, Shinya H, Hiroaki H. Determination of the optimal method for the field-in-field technique in breast tangential radiotherapy. J Radiat Res (Tokyo) (2014) 4):769–73. 10.1093/jrr/rrt233 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Donovan E, Bleakley N, Denholm E, Evans P, Gothard L, Hanson J, et al. Randomised trial of standard 2D radiotherapy (RT) versus intensity modulated radiotherapy (IMRT) in patients prescribed breast radiotherapy. Radiother Oncol (2007) 82(3):254–64. 10.1016/j.radonc.2006.12.008 [DOI] [PubMed] [Google Scholar]
- 18. Ho AY, Mccormick B. A Multicenter Randomized Trial of Breast Intensity-Modulated Radiation Therapy to Reduce Acute Radiation Dermatitis. J Clin Oncol (2009) 20(1):89–90. 10.1016/S1043-321X(09)79202-1 [DOI] [PubMed] [Google Scholar]
- 19. Fong A, Bromley R, Beat M, Vien D, Morgan G. Dosimetric comparison of intensity modulated radiotherapy techniques and standard wedged tangents for whole breast radiotherapy. J Med Imaging Radiat Oncol (2009) 53(1):92–9. 10.1111/j.1754-9485.2009.02043.x [DOI] [PubMed] [Google Scholar]
- 20. Schubert LK, Gondi V, Sengbusch E, Westerly DC, Soisson ET, Paliwal BR, et al. Dosimetric comparison of left-sided whole breast irradiation with 3DCRT, forward-planned IMRT, inverse-planned IMRT, helical tomotherapy, and topotherapy. Radiother Oncol (2011) 100(2):241–6. 10.1016/j.radonc.2011.01.004 [DOI] [PubMed] [Google Scholar]
- 21. Osman SO, Hol S, Poortmans PM, Essers M. Volumetric modulated arc therapy and breath-hold in image-guided locoregional left-sided breast irradiation. Radiother Oncol (2014) 112(1):17–22. 10.1016/j.radonc.2014.04.004 [DOI] [PubMed] [Google Scholar]
- 22. Shanmugam Thirumalai S, Chandrasekaran Anu R, Murugesan K, Gandhi A, Shanmuga S. Feasibility study of deep inspiration breath-hold based volumetric modulated arc therapy for locally advanced left sided breast cancer patients. Asian Pac J Cancer Prev (2014) 15(20):9033–8. 10.7314/APJCP.2014.15.20.9033 [DOI] [PubMed] [Google Scholar]
- 23. Yang D, Yoon W, Chung S, Lee J, Lee S, Park Y, et al. Set-up uncertainty during breast radiotherapy. Strahlenther Onkol (2013) 189:315–20. 10.1007/s00066-012-0271-4 [DOI] [PubMed] [Google Scholar]
- 24. Kron T, Lee C, Perera F, Yu E. Evaluation of intra- and inter-fraction motion in breast radiotherapy using electronic portal cine imaging. Technol Cancer Res Treat (2004) 3(5):443–9. 10.1177/153303460400300505 [DOI] [PubMed] [Google Scholar]
- 25. Alaei P, Spezi E. Imaging dose from cone beam computed tomography in radiation therapy. Phys Med (2015) 31(7):647–58. 10.1016/j.ejmp.2015.06.003 [DOI] [PubMed] [Google Scholar]
- 26. Batumalai V, Quinn A, Jameson M, Delaney G, Holloway L. Imaging dose in breast radiotherapy: does breast size affect the dose to the organs at risk and the risk of secondary cancer to the contralateral breast? J Med Radiat Sci (2015) 62(1):32–9. 10.1002/jmrs.91 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27. Bartlett FR, Colgan RM, Donovan EM, Carr K, Landeg S, Clements N, et al. Voluntary breath-hold technique for reducing heart dose in left breast radiotherapy. J Vis Exp (2014) 89(89):e51578. 10.3791/51578 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28. Bartlett FR, Colgan RM, Carr K, Donovan EM, Mcnair HA, Locke I, et al. The UK HeartSpare study: randomised evaluation of voluntary deep-inspiratory breath-hold in women undergoing breast radiotherapy. Radiother Oncol (2013) 108(2):242–7. 10.1016/j.radonc.2013.04.021 [DOI] [PubMed] [Google Scholar]
- 29. White JT, Arthur D, Buchholz T, MacDonald S, Marks L, Pierce L, et al. Breast Cancer Atlas for Radiation Therapy Planning: Consensus Definitions. RTOG; (2017). Available at: http://www.rtog.org/. [Google Scholar]
- 30. Apar G, Nisha O, Haffty BG. Hypofractionated radiation treatment in the management of breast cancer. Expert Rev Anticancer Ther (2018) 18(8):793–803. 10.1080/14737140.2018.1489245 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31. Whelan T. Randomized Trial of Breast Irradiation Schedules After Lumpectomy for Women With Lymph Node-Negative Breast Cancer. Cancer Spectr Knowl Environ (2002) 94(15):1143–50. 10.1093/jnci/94.15.1143 [DOI] [PubMed] [Google Scholar]
- 32. Whelan TJ, Pignol J-P, Levinel MN, Julian JA, MacKenzie R, Parpia S, et al. Long-Term Results of Hypofractionated Radiation Therapy for Breast Cancer. N Engl J Med (2010) 362(6):513–20. 10.1056/NEJMoa0906260 [DOI] [PubMed] [Google Scholar]
- 33. Marks LB, Yorke ED, Jackson A, Haken RKT, Constine LS, Eisbruch A, et al. Use of Normal Tissue Complication Probability Models in the Clinic. Int J Radiat Oncol Biol Phys (2010) 76(3 Suppl):S10. 10.1016/j.ijrobp.2009.07.1754 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34. Haken RKT, Marks LB, Bentzen SM, Constine LS, Deasy JO, Eisbruch A, et al. Quantitative Analyses of Normal Tissue Effects in the Clinic (QUANTEC): Clinical Use. Radiother Oncol (2009) 98(Suppl 2):S8–S. 10.1007/978-3-642-03474-9_302 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35. Lai Y, Chen Y, Wu S, Shi L, Fu L, Ha H, et al. Modified Volumetric Modulated Arc Therapy in Left Sided Breast Cancer After Radical Mastectomy With Flattening Filter Free Versus Flattened Beams. Medicine (2016) 95(14):e3295. 10.1097/MD.0000000000003295 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36. Tamburella C, Faessler G, Guibert G, Hirschi L, Pion L, Weber P, et al. EP-1953: Dosimetric effect of using FFF-VMAT beams for left-sided breast radiotherapy in breath hold. Radiother Oncol (2018) 127:S1061–S2. 10.1016/S0167-8140(18)32262-X [DOI] [Google Scholar]
- 37. Spruijt KH, Dahele M, Cuijpers JP, Jeulink M, Rietveld D, Slotman BJ, et al. Flattening Filter Free vs Flattened Beams for Breast Irradiation. Int J Radiat Oncol Biol Phys (2013) 85(2):506–13. 10.1016/j.ijrobp.2012.03.040 [DOI] [PubMed] [Google Scholar]
- 38. Finazzi T, Nguyen VT, Zimmermann F, Papachristofilou A. Impact of patient and treatment characteristics on heart and lung dose in adjuvant radiotherapy for left-sided breast cancer. Radiat Oncol (2019) 14(1):153. 10.1186/s13014-019-1364-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39. Lazzari G, Terlizzi A, Leo MG, Silvano G. VMAT radiation-induced nausea and vomiting in adjuvant breast cancer radiotherapy: The incidental effect of low-dose bath exposure. Clin Trans Radiat Oncol (2017) 7:43–8. 10.1016/j.ctro.2017.09.009 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40. Borges C, Cunha G, Monteiro-Grillo I, Vaz P, Teixeira N. Comparison of different breast planning techniques and algorithms for radiation therapy treatment. Phys Med (2014) 30:160–70. 10.1016/j.ejmp.2013.04.006 [DOI] [PubMed] [Google Scholar]
- 41. Darby S, Mcgale P, Correa C, Taylor T, Arriagada R, Clarke M, et al. Effect of radiotherapy after breast-conserving surgery on 10-year recurrence and 15-year breast cancer death: meta-analysis of individual patient data for 10,801 women in 17 randomised trials. Lancet (2011) 378(9804):1707–16. 10.1016/S0140-6736(11)61629-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42. Mcgale P, Taylor C, Correa C, Cutter D, Duane F, Ewertz M, et al. Early Breast Cancer Trialists Collaborative Group (EBCTCG). Effect of radiotherapy after mastectomy and axillary surgery on 10-year recurrence and 20-year breast cancer mortality: meta-analysis of individual patient data for 8135 women in 22 randomised trials. Lancet (2014) 383(9935):2127–35. 10.1016/S0140-6736(14)60488-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43. Carolyn T, Candace C, Duane FK, Aznar MC, Anderson SJ, Jonas B, et al. Estimating the Risks of Breast Cancer Radiotherapy: Evidence From Modern Radiation Doses to the Lungs and Heart and From Previous Randomized Trials. J Clin Oncol (2017) 35(15):1641–9. 10.1200/JCO.2016.72.0722 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44. Darby S, Davies C, McGale P. The Early Breast Cancer Trialists Collaborative Group: a brief history of results to date. In: Davison AC, Dodge Y, Wermuth N, editors. Celebrating Statistics. Oxford: Oxford University Press; (2005). p. 185–97. [Google Scholar]
- 45. Killander F, Wieslander E, Karlsson P, Holmberg E, Lundstedt D, Holmberg L, et al. No Increased Cardiac Mortality or Morbidity of Radiation Therapy in Breast Cancer Patients After Breast-Conserving Surgery: 20-Year Follow-up of the Randomized SweBCGRT Trial. Int J Radiat Oncol Biol Phys (2020) 107(4):701–9. 10.1016/j.ijrobp.2020.04.003 [DOI] [PubMed] [Google Scholar]
- 46. Abo-Madyan Y, Aziz MH, Aly MMOM, Schneider F, Sperk E, Clausen S, et al. Second cancer risk after 3D-CRT, IMRT and VMAT for breast cancer. Radiother Oncol (2014) 110(3):471–6. 10.1016/j.radonc.2013.12.002 [DOI] [PubMed] [Google Scholar]
- 47. Jensen CA, Roa AMA, Johansen M, Lund JÅ, Frengen J. Robustness of VMAT and 3DCRT plans toward setup errors in radiation therapy of locally advanced left-sided breast cancer with DIBH. Phys Med (2018) 45:12–8. 10.1016/j.ejmp.2017.11.019 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.