Table 1.
Comparison of signals acquired using textile and gelled electrodes, mean and SD over 25 consecutive cycles.
Waveforms and Intervals | Electrode Type | ||
---|---|---|---|
Silver Printed Cotton | Silver Printed Polyester | Ag/AgCl | |
P (mV) | 0.09 ± 0.01 | 0.09 ± 0.02 | 0.09 ± 0.01 |
R-peak amplitude (mV) | 1.28 ± 0.05 | 1.26 ± 0.04 | 1.29 ± 0.05 |
T (mV) | 0.35 ± 0.04 | 0.35 ± 03 | 0.35 ± 0.03 |
PR (ms) | 158 ± 4 | 154 ± 4 | 158 ± 4 |
QRS (ms) | 93 ± 1 | 93 ± 1 | 93 ± 1 |
QT (ms) | 346 ± 8 | 336 ± 8 | 348 ± 8 |
SNR (dB) | 33.10 ± 1.31 | 30.17 ± 1.43 | 33.52 ± 1.30 |
Hear rate (bpm) | 75 ± 1 | 73 ± 2 | 76 ± 1 |