HR and R-R interval of textile and gelled electrodes of each subject, mean and SD over 25 consecutive cycles.

Subject HR (bpm) R-R Interval (ms)
Silver Printed Cotton Silver Printed Polyester Ag/AgCl Silver Printed Cotton Silver Printed Polyester Ag/AgCl
1 73 75 76 736 ± 18 756 ± 20 762 ± 16
2 80 79 79 760 ± 23 712 ± 40 736 ± 16
3 82 84 82 761 ± 27 720 ± 27 727 ± 39
4 80 80 82 763 ± 20 768 ± 28 736 ± 20
5 90 92 93 643 ± 12 653 ± 14 641 ± 15
6 78 77 77 797 ± 49 782 ± 53 807 ± 45