Biomarker Sensing Platform Detection Method LODa Detection Range Recoveryb Validation Assay Time Sample Volume Preparation Reference
CYFRA-21–1 BSA/anti-CYFRA-21–1/APTES/ZrO2-rGO/ITO/glass DPV 0.122 ng mL−1 2 – 22 ng mL−1 N/A ELISA RSD: 0.2– 3.35% 16 min N/A Whole saliva (Kumar et al. 2016)
CYFRA-21–1 BSA/anti-CYFRA-21–1/APTES/nHfO2@RGO/ITO/glass DPV 0.16 ng mL−1 0–30 ng mL−1 N/A ELISARSD: 2.2–8% 15 mn N/A N/A (Kumar et al. 2018)
Interleukin-6 (IL-6) Aptameric GFET Potentio-metric 12.2 Pm 0.05 – 0.8 nM N/A N/A 400s 30 μL 1:4 dilution with PBS (Hao et al. 2019)
Interleukin-8 (IL-8) AuNPs-rGO/ITO/glass DPV 73 pg mL−1 500 fg mL−1 – 4 ng mL−1 94.15% N/A 9 min N/A Whole saliva (Verma et al. 2017)
Interleukin-8 (IL-8) Anti-IL-8/ZnO-rGO/ITO/glass DPV 51.53 pg mL−1 100 fg/mL – 5 ng mL−1 95 – 100% N/A 10 min N/A 1:10 dilution with 1 × PBS (Verma and Singh 2019)
Prostate specific antigen (PSA) Graphene-polymer-PS67-b-PAA27-Au Chemi-resistive 40 fg mL−1 100 fg/mL – 100 ng mL−1 94 – 114% Self-designed ELISA protocol: 94.02% linearity 1–3 min 10 μL Whole saliva (Khan et al. 2018)
Prostate specific antigen (PSA) MWCNT/His-rGO/thionine/Ab/GCE DPV 2.8 fg mL−1 10 fg mL−1 – 20 ng mL−1 96.4 – 101.9% ELISA 97.6% accuracy 25 min N/A 1:10 dilution with PBS (Farzin et al. 2019)
3-Nitro-l-tyrosine (3-NT) ZrO2@rGONCs/GCE Ampero-metric 0.009 μM 0.025 – 855.2 μM 94.61 – 96.24% N/A N/A N/A Whole saliva (Maheshwaran et al. 2020)
8-OHdG Graphene modified SPE CV 1.8 nM 6×10−6 – 6×10−4 M 95 – 106% N/A N/A N/A Whole saliva (Varodi et al. 2019)
a

S/N = 3, otherwise differently specified

b

values obtained in real saliva samples, otherwise differently specified