Table 3.
Comparison of reliability and validity of smartphone and visual estimation measurements against universal goniometer measurements within observers (intraobserver reliability)
Movement / method | ICC (95% CI) | Mean absolute difference ± sd (°) | Bland–Altman mean difference (95% CI) (°) | ± 95% LOA (°) | P value* |
---|---|---|---|---|---|
Flexion | |||||
Visual estimation | 0.92 (0.77–0.96) | 7.7 ± 5.9 | 4.5 (3.0–5.9)3.04–5.89 | ± 12.5 | < 0.001 |
Smartphone | 0.98 (0.97–0.99) | 2.6 ± 1.9 | 0.3 –(–0.4–0.9)-0.43–0.95 | ± 6.1 | 0.450 |
Extension | |||||
Visual estimation | 0.98 (0.97–0.99) | 2.9 ± 3.3 | 0.7 (–0.1–1.4)-0.13–1.43 | ± 3.5 | 0.102 |
Smartphone | 0.99 (0.98–0.99) | 1.1 ± 1.0 | –0.2 (–0.3–0.3)-0.32–0.27 | ± 2.6 | 0.867 |
Supination | |||||
Visual estimation | 0.97 (0.96–0.98) | 3.9 ± 3.3 | 0.5 (–0.4–1.4)-0.44–1.44 | ± 8.3 | 0.295 |
Smartphone | 0.99 (0.98–0.99) | 2.2 ± 1.5 | 0.7 (0.2–1.2)0.19–1.21 | ± 4.5 | 0.108 |
Pronation | |||||
Visual estimation | 0.91 (0.86–0.94) | 4.8 ± 3.5 | 0.4 (–0.9–1.7)-0.95–1.67 | ± 11.5 | 0.008 |
Smartphone | 0.98 (0.97–0.99) | 1.9 ± 1.1 | –0.1 (–0.6–0.4)-0.64–0.37 | ± 4.4 | 0.589 |
CI, confidence interval; ICC, intraclass correlation coefficient; LOA, limits of agreement
P values calculated via paired-T test for mean absolute differences. Significance set at α = 0.05