Table 1. Commercially available intravascular imaging modalities.
| Imaging modality | Manufacturer | Source | Axial resolution (µm) | Lateral resolution (µm) | Tissue penetration (mm) | Frame width (mm) | Pullback distance (mm) | 
|---|---|---|---|---|---|---|---|
| Intravascular ultrasound | Boston Scientific | 40 MHz Ultrasound | 38 | 80–200 | >5 | 0.02–0.03 | 100 | 
| Volcano | 20 MHz Ultrasound | 170 | 100 | >5 | 0.02–0.03 | 150 | |
| Volcano | 45 MHz Ultrasound | 46 | 179 | >5 | 0.02–0.03 | 150 | |
| InfraReDx/Nipro | 50 MHz Ultrasound | 20 | 240–260 | 8 | 0.02–0.07 | 150 | |
| Boston Scientific | 60 MHz Ultrasound | 22 | 50–140 | >5 | 0.02–0.03 | 100 | |
| ACIST | 60 MHz Ultrasound | 40 | 90 | 3 | 0.02–0.17 | 120 | |
| Terumo | 60 MHz Ultrasound | <30 | 100 | Not available | 0.02–0.1 | 150 | |
| Optical coherence tomography | Abbott | Near Infrared 1.3 ìm | 15 | 20–40 | 1 to 2 | 0.1–0.2 | 75 | 
| Terumo | Near Infrared | <20 | 30 | Not available | 0.13–0.25 | 150 |