Skip to main content
. Author manuscript; available in PMC: 2021 Dec 1.
Published in final edited form as: Anesth Analg. 2020 Dec;131(6):1923–1933. doi: 10.1213/ANE.0000000000005220

Table 2:

Disassembly and commentary for the TMS320C32 DSP BIS Engine routines that calculate total signal power in decibels (TOTPOW) and 95th percentile spectral edge frequency in Hz (SEF). A small bug, shown in boldface, is present at 8FA72C, which causes nonsensical spectral edge frequencies to be returned if there is no signal power present (i.e. a perfectly isoelectric EEG).

graphic file with name nihms-1638214-t0006.jpg