TABLE 4.

Reasons and frequencies of unsuccessful MBW measurements

Finding in individual measurements Unsuccessful MBW investigations, n=16 Successful MBW investigations with at least 1 unacceptable measurement, n=55
Leak 5 (31.3) 27 (49.1)
End of test criteria not met 2 (12.5) 10 (18.2)
Irregular breathing pattern 2 (12.5) 5 (9.1)
End of test criteria not met/irregular breathing pattern 2 (12.5) 4 (7.3)
Leak/end of test criteria not met 1 (6.3) 3 (5.5)
Leak/end of test criteria not met / irregular breathing pattern 2 (12.5) 1 (1.8)
End of test criteria not met/no washout started 0 (0.0) 2 (3.6)
Leak/irregular breathing pattern 1 (6.3) 0 (0.0)
Leak/swallow at critical time point 0 (0.0) 1 (1.8)
Leak/no washout started 0 (0.0) 1 (1.8)
No washout started 0 (0.0) 1 (1.8)
Use of wrong DSR set 1 (6.3) 0 (0.0)

Data are presented as n (%). MBW: multiple-breath washout; DSR: dead space reducer.