. 2020 Nov 23;10:20407. doi: 10.1038/s41598-020-76989-0

Table 2.

The number of reproducible features for intra-MR analysis with different acquisition parameters for volunteers and kiwis out of a total of 396.

Group n Number of reproducible features (%) Mean ± standard deviation
CV < 0.1 CV < 0.15
AM
Kiwis G.0 3 290(73.2%) 324(81.8%) 0.12 ± 0.29
Volunteers G.1 3 186(47.0%) 231(58.3%) 0.31 ± 1.41
G.2 3 166(41.9%) 196(49.5%) 0.23 ± 0.65
G.3 3 185(46.7%) 221(55.8%) 0.25 ± 0.47
ST
Kiwis G.0 3 324(81.85) 351(88.6%) 0.09 ± 0.19
Volunteers G.1 3 193(48.7%) 236(59.6%) 0.24 ± 0.66
G.2 3 190(48.0%) 239(60.4%) 0.18 ± 0.24
G.3 3 197(49.7%) 238(60.1%) 0.23 ± 0.83
TE
Kiwis G.0 3 212(53.5%) 329(83.1%) 0.14 ± 0.22
Volunteers G.1 3 172(43.4%) 227(57.3%) 0.32 ± 0.74
G.2 3 154(38.9%) 204(51.5%) 0.24 ± 0.33
G.3 3 183(46.2%) 208(52.5%) 0.23 ± 0.35
TR
Kiwis G.0 3 356(89.9%) 362(91.4%) 0.08 ± 0.23
Volunteers G.1 3 147(37.1%) 219(55.3%) 0.27 ± 0.71
G.2 3 194(49.0%) 214(54.0%) 0.22 ± 0.72
G.3 3 157(39.6%) 201(50.7%) 0.35 ± 0.73

TR repetition time; TE echo time; ST slice thickness; AM acquisition matrix; G.0 represents the three kiwis, while G.1–G.3 represent the three group volunteers in 6th–10th, 11th–15th, and16th–20th day of physiological cycle respectively; The values displayed on the table were means within each group. “Mean ± Standard Deviation” was calculated from mean CV values within each group. MR scanner: Philips Medical Systems (Ingenia 3.0 T, Philips Healthcare, Best, The Netherlands).