Skip to main content
. 2020 Aug 21;12(4):383–392. doi: 10.5114/jcb.2020.98119

Table 1.

Positional repeatability based on five CT scans, with the SPS source at each chosen dwell position for the VeneziaTM lunar ovoids. 1300 mm represents the most distal treatment dwell position

1300 mm (distal) 1290 mm 1280 mm
22 mm Maximum 3D difference (mm) 0.3 0.2 0.2
Average 3D difference (mm) 0.2 0.1 0.1
Standard deviation (mm) 0.1 0.1 0.1
26 mm Maximum 3D difference (mm) 0.5 0.2 0.2
Average 3D difference (mm) 0.2 0.1 0.1
Standard deviation (mm) 0.2 0.1 0.1
30 mm Maximum 3D difference (mm) 0.6 0.3 0.3
Average 3D difference (mm) 0.3 0.2 0.1
Standard deviation (mm) 0.2 0.1 0.1