. 2020 Aug 21;12(4):383–392. doi: 10.5114/jcb.2020.98119

Table 2.

Flexitron® 192Ir source position on radiochromic film versus SPS centroid in the split-ring plane from CT scans of the VeneziaTM lunar ovoids. Positive values denote that the 192Ir source is more distal than the SPS

1300 mm (distal) 1290 mm 1280 mm
22 mm Maximum 2D difference (mm) 0.4 0.4 0.3
Average 2D difference (mm) 0.2 0.2 0.2
Standard deviation (mm) 0.2 0.2 0.1
26 mm Maximum 2D difference (mm) 0.7 0.5 0.6
Average 2D difference (mm) 0.6 0.3 0.3
Standard deviation (mm) 0.1 0.2 0.3
30 mm Maximum 2D difference (mm) 0.4 0.7 0.3
Average 2D difference (mm) 0.3 0.5 0.2
Standard deviation (mm) 0.2 0.3 0.1