Table 3.
Type A (mm) | Type B (mm) | |
---|---|---|
SPS repeatability from CT scans | 0.1 | – |
SPS agreement with 192Ir source | 0.3 | – |
Catheter reconstruction method | – | 0.1 |
CT registration accuracy | – | 0.4 |
Combined uncertainty (k = 1) | 0.5 | |
Expanded uncertainty (k = 2) | 1.0 |
SPS – source position simulator