Skip to main content
. 2020 Aug 21;12(4):383–392. doi: 10.5114/jcb.2020.98119

Table 5.

Tissue volumes with dose differences exceeding ±5% and ±10% of the prescription dose using different catheter reconstruction techniques relative to the SPS-based plan with three VeneziaTM lunar ovoid sizes

22 mm 26 mm 30 mm
V–5% | V+5%
(cm3)
V–10% | V+10%
(cm3)
V–5% | V+5%
(cm3)
V–10% | V+10%
(cm3)
V–5% | V+5%
(cm3)
V–10% | V+10%
(cm3)
Centerline 1.75 | 1.15 0.71 | 0.33 4.89 | 3.04 1.94 | 1.18 6.34 | 4.77 2.50 | 2.03
CT wire 0.76 | 1.54 0.27 | 0.62 4.00 | 3.94 1.54 | 1.67 3.25 | 2.59 1.28 | 1.01
Manufacturer’s model 1.12 | 1.26 0.21 | 0.34 1.19 | 1.88 0.31 | 0.80 2.54 | 1.60 0.67 | 0.33