Skip to main content
. 2020 Aug 21;12(4):383–392. doi: 10.5114/jcb.2020.98119

Table 6.

Maximal dose differences for tissue volumes of 0.1 cm3, 1.0 cm3, and 2.0 cm3 given as a percentage of the prescription dose using different catheter reconstruction techniques relative to the SPS-based plan with three VeneziaTM lunar ovoid sizes

22 mm 26 mm 30 mm
ΔD0.1cm3
(%)
ΔD1cm3
(%)
ΔD2cm3
(%)
ΔD0.1cm3
(%)
ΔD1cm3
(%)
ΔD2cm3
(%)
ΔD0.1cm3
(%)
ΔD1cm3
(%)
ΔD2cm3
(%)
Centerline 19 5.3 3.8 37 10 6.9 33 10 7.6
CT wire 28 7.5 4.2 39 15 9.1 35 9.6 5.6
Manufacturer’s model 21 5.5 3.6 28 8.2 4.4 16 5.8 4.4