Skip to main content
. 2020 Aug 21;12(4):383–392. doi: 10.5114/jcb.2020.98119

Table 7.

Differences in DVH parameters for different catheter reconstruction techniques relative to the SPS- based plan for clinical VeneziaTM plans without interstitial needles (P1) and with interstitial needles (P2)

Centerline (%) CT wire (%) Manufacturer’s model (%)
P1 P2 P1 P2 P1 P2
HR-CTV Dmin 1.5 5.3 –4.0 0.8 1.8 1.6
D90% –0.8 –0.4 –1.0 0.5 –0.8 0.5
Bladder D2cm3 –0.8 2.8 –0.2 0.2 –0.2 0.0
Bowel D2cm3 –1.6 –2.7 0.0 0.0 0.5 –2.0
Rectum D2cm3 –1.4 –2.3 –0.5 –0.3 –1.8 –0.8
Sigmoid D2cm3 3.6 –3.4 0.0 0.0 2.4 –0.7
Body V200% –2.0 –3.9 2.0 1.5 2.0 –1.5
Vagina PIBS+2 cm –0.7 –0.2 –0.4 1.1 –0.3 –0.1
PIBS –1.3 –1.0 0.1 1.4 –0.6 0.6
PIBS–2 cm –0.2 –0.1 –0.5 2.2 –0.2 –0.2
Dsurf –4.7 3.1 –2.1 3.1 –4.4 1.9
D5mm –3.6 1.3 –0.7 1.4 –2.8 0.8