Skip to main content
. 2020 Aug 23;10(2):020425. doi: 10.7189/jgh.10.020425

Table 1.

Power storage (PS) system – oxygen concentration output at maximum flow (5 L/min), and availability of oxygen for clinical care

Outcome measured Number/proportion
Oxygen concentration
Recordings (No.)
56

Oxygen concentration output, Median (range)
92.5% (82.2-95.4)

Proportion of recordings ≥82% ± 3% concentration
100%
Availability of oxygen to patients:

Pulse oximetry readings (No.)
12863

Pulse oximetry readings <90% SpO2 (No.)
1156
Pulse oximetry readings <90% for which oxygen was available, No. (%)* 1156 (100%)

SpO2 – peripheral capillary oxygen saturation