Solar-power storage (PS) system – oxygen concentration output at maximum flow (5 L/min), and availability of oxygen for clinical care

Oxygen concentration Outcome measured Number/Proportion

Recordings (No.)
38

Oxygen concentration output, Median (range)
89.8% (88.0-99.7)

Proportion of recordings ≥82% ± 3% concentration
100%
Availability of oxygen to patients:

Pulse oximetry readings (N)
13 862

Pulse oximetry readings <90% SpO2 (No.)
1974
Pulse oximetry readings <90% for which oxygen was available, No. (%)* 1974 (100%)

SpO2 – peripheral capillary oxygen saturation