Summary of readings from all four sensors (A, B, C and D) from four different positions on the left wrist of healthy volunteer P9; non-pulsatile component (DC) level (nW), the average PPG peak to peak (P-Pave), Perfusion Index (PI) average and SNR for both wavelengths; and the average SpO2 levels from both the plastic optical fibre (POF) and Masimo sensors.

Sensor Signal DC Level
(nW)
P-Pave (nW) PI
(%)
SNR Average
SpO2 (%)
POF Masimo
A Red 502.8 0.20 0.04 6.63 98.9
IR 186.0
B Red 420.7 0.84 0.19 7.19 98.1 98.3
IR 191.3 0.15 0.08 4.17
C Red 475.8 99
IR 188.1
D Red 408.9 0.52 0.13 5.98 99.5 99
IR 178.1 0.10 0.06 4.29