Published in final edited form as: Med Phys. 2019 Oct 31;46(12):5514–5527. doi: 10.1002/mp.13862

Table 1:

Seven sets of CT parameters used in the CT examinations. All sets used the same common settings for clinical coronary protocol stated in the body of the text. Only the listed parameters in the table columns were varied among 7 sets. All reconstructions were performed with the standard GE reconstruction software with ASiR-V=50%. Focal spot size = 1.0 × 0.7mm for Small, and 1.6 × 1.2mm for Large, and 2.0 × 2.1mm for Extra Large.

Set No. Tube energy (kV) Tube current (mA) focal spot size (mm) Reconstruction kernel CTDIvol (mGy) Marker in the plot
1 100 640 Large(L) Standard(std) 4.69
2 100 640 Large(L) Bone 4.69 +
3 100 645 Extra Large(XL) Bone 4.60 *
4 100 420 Small(S) Bone 3.11 ×
5 120 350 Small(S) Standard(std) 4.12
6 100 420 Small(S) Standard(std) 3.11
7 80 580 Small(S) Standard(std) 2.11