Tolerances from TG‐142 and machine specifications for the Varian TrueBeam® for all cone‐beam computed tomography (CBCT) image quality metrics.

Parameter TG −142 Varian TrueBeam®
Geometric distortion

<2 mm for non‐SRS/SBRT

<1 mm for SRS/SBRT

N/A
Spatial resolution (full fan) Baseline ≥6 lp/mm
Spatial resolution (half fan) Baseline ≥4 lp/mm
HU uniformity Baseline ±40 HU
Contrast Baseline N/A
Noise Baseline N/A
HU accuracy Baseline ±50 HU