. 2020 Oct 19;21(11):215–225. doi: 10.1002/acm2.13062

Table 5.

Measurements of uniformity for cone‐beam computed tomographies (CBCTs) acquired immediately after a high‐dose scan and CBCTs acquired with at least a 10‐min interval between scans for the Head and Pelvis techniques.

Scan number Post high‐dose scan 10‐min interval
Head Pelvis Head Pelvis
1 −79.34 −30.68 −3.48 −17.02
2 −41.93 −43.05 −3.68 −16.31
3 −55.66 −48.39 7.29 −20.88
Average −58.98 −40.71 −4.05 −18.07