Table 3.

Comparisons of mean gamma passing rates (across the five cross-validation folds) for low-resolution dose (with 3 mm or 5 mm grids) and predicted high-resolution dose in the planning target volume (PTV), bladder, and rectum.

ROI Gamma passing rate with a 2%/2 mm criterion (%)
AAA AXB
3 mm grid Prediction AAA 3 model 5 mm grid Prediction AAA 5 model 3 mm grid Prediction AXB 3 model
PTV 93.6 (3.7) 99.1 (1.0) 71.0 (5.8) 92.4 (4.6) 88.0 (6.2) 98.2 (2.6)
Bladder 98.1 (2.1) 99.5 (1.3) 90.0 (4.7) 97.9 (2.3) 98.0 (2.1) 99.1 (2.4)
Rectum 98.8 (0.8) 100.0 (0.1) 72.4 (6.9) 99.0 (1.1) 98.7 (0.6) 99.4 (0.4)

The standard deviations are indicated in parentheses.

ROI, region of interest; AAA, anisotropic analytical algorithm; AXB, Acuros XB.